Beckman Coulter Inc. recalls
28 recalls filed under Beckman Coulter Inc.. The most recent was published October 14, 2026.
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE recall: Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of…
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE recall: Beckman Coulter has identified a software issue affecting DxC 500i Clinical
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a…
-
Class II · 2 months ago
Beckman Coulter DxC 500i Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23352, MicroScan LabPro V5.03 Panel Update 11 recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
-
Class II · 6 months ago
Beckman Coulter, Inc. Access Total T4 Calibrator recall: Beckman Coulter has identified that Access Total T4 Calibrator lots (PN
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes…
-
Class II · 6 months ago
Beckman Coulter, Inc. MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code recall: Due to manufacturing issue (incorrectly processing) with in vitro multidrug
Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
-
Class II · 6 months ago
Beckman Coulter, Inc. MicroScan Neg MIC 3J REF C54814 recall: Due to a manufacturing issue - missing a drug in well 4/4 but contain double
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
-
Class II · 7 months ago
Beckman Coulter, Inc. MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 recall: Due to the likely presence of contamination in well(s)
Due to the likely presence of contamination in well(s).
-
Class II · 9 months ago
Beckman Coulter, Inc. CellMek SPS recall: Sample preparation system designed to automate staining, lysing
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or…
-
Class II · 9 months ago
Beckman Coulter, Inc. Coulter DxH Diluent, REF 628017 Coulter DxH ECO Diluent REF recall: Due to specific lots of diluents contributing to elevated platelet Daily
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
-
Class II · 9 months ago
Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456 recall: Beckman Coulter has determined that, due to a potential communication issue
Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a…
-
Class II · 9 months ago
Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260 recall: Beckman Coulter has determined that, due to a potential communication issue
Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a…
-
Class II · 10 months ago
Beckman Coulter, Inc. Access 2 Reaction Vessels, individual, disposable recall: It has been determined that certain lots of Access 2 Reaction Vessels
It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction…
-
Class II · 11 months ago
Beckman Coulter Inc. DxFLEX Flow Cytometer recall: Due to a software error code that crashes during acquisition on patient
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to…
-
Class II · 12 months ago
Beckman Coulter Inc. Synchron Systems Direct LDL Cholesterol Reagent, REF: 969706 recall: Beckman Coulter has identified that there is an increased probability
Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166…
-
Class II · 12 months ago
Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456 recall: An internal investigation has determined that due to defective sub-components
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may…
-
Class II · 12 months ago
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260 recall: An internal investigation has determined that due to defective sub-components
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may…
-
Class II · 12 months ago
Beckman Coulter, Inc. CellMek SPS Sample Preparation System recall: Automated pipetting, diluting and specimen processing workstations for flow
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead…
-
Class II · 12 months ago
Beckman Coulter, Inc. DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE recall: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct…
-
Class II · 12 months ago
Beckman Coulter, Inc. DxI 9000 Access Immunoassay Analyzer recall: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct…
-
Class II · 1+ years ago
Beckman Coulter Inc. Synchron Systems Phosphorus Reagent REF: 467868 For In Vitro recall: Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868)
Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which…
-
Class II · 1+ years ago
Beckman Coulter Inc. DxC 500 AU Module w/ISE, DxC 500i, With ISE recall: On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination
On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is…
About this page
Beckman Coulter Inc. appears here because a federal agency named it as the recalling firm or the brand on the label. A recall is not a judgement about everything the company sells — it covers only the lot codes and models listed in each notice.