Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260 recall: An internal investigation has determined that due to defective sub-components
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0428-2026
- Published
- November 12, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Beckman Coulter, Inc.
- Product type
- Medical devices
- Units affected
- 80 units
- Made in
- United States
- Lot / code info
- A71460, UDI/DI 15099590369224, Serial Numbers: 902515, 902529, 902528, 902534, 902526, 902527, 902523, 902522, 902524, 902505, 902578, 902579, 902545, 902530, 902531, 902581, 902504, 902514, 902525, 902535, 902566, 902516, 902518, 902532, 902555, 902558, 902564, 902519, 902521, 902559, 902511, 902549, 902554, 902547, 902582, 902583; A30260, UDI/DI 15099590341602, Serial Numbers: 902536, 902541, 902517, 902563, 902565, 902557, 902513, 902507, 902543, 902560, 902562, 902506, 902508, 902574, 902533, 902572, 902540, 902569, 902568, 902577, 902510, 902580, 902567, 902573, 902570, 902571, 902538, 902553, 902552, 902512, 902539, 902561, 902575, 902544, 902542, 902556, 902537, 902509, 902520, 902546, 902548, 902551, 902550, 902576.
- UPC codes
- 15099590369224, 15099590341602
Products covered by this recall
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Check your product
The notice names 14 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 15099590369224
- 15099590341602
Model or item number
On the rating label, the box or the instruction sheet.
- A71460
- A30260
Serial number
On the rating plate, often next to the date of manufacture.
- 902515
- 902529
- 902528
- 902534
- 902526
- 902536
- 902541
- 902517
- 902563
- 902565
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Beckman Coulter, Inc. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Beckman Coulter, Inc. — 1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Beckman Coulter, Inc. medical devices recalled?
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results. An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 15099590369224, 15099590341602; model number A71460, A30260. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Related recalls
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE recall: Beckman Coulter has identified a software issue affecting DxC 500i Clinical
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a…
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE recall: Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of…
-
Class II · 2 months ago
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter DxC 500i Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23352, MicroScan LabPro V5.03 Panel Update 11 recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."