Beckman Coulter, Inc. DxI 9000 Access Immunoassay Analyzer recall: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
The hazard: Device malfunction
This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
What the notice says: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.
Symptoms and when to call a doctor
If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-0216-2026
- Published
- October 29, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Beckman Coulter, Inc.
- Product type
- Medical devices
- Units affected
- 571 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- C11137 UDI-DI: 15099590732103; Serial Numbers: 300621 300655 300564 300572 300411 300425 300576 300577 300625 300295 300713 300714 300289 300497 300333 300520 300524 300440 300443 300441 300444 300445 300528 300515 300332 300466 300484 300412 300395 300541 300255 300270 300499 300435 300436 300467 300532 300560 300600 300604 300605 300607 300609 300610 300707 300709 300710 300596 300597 300374 300679 300677 300558 300559 300593 300584 300169 300239 300124 300225 300595 300638 300495 300496 300381 300668 300237 300156 300571 300641 300131 300637 300488 300142 300141 300143 300151 300152 300316 300546 300298 300299 300328 300414 300413 300422 300132 300534 300317 300318 300567 300568 300163 300162 300276 300277 300174 300226 300370 300459 300386 300503 300388 300389 300745 300742 300204 300216 300247 300246 300438 300149 300662 300716 300268 300380 300391 300429 300437 300303 300305 300349 300353 300290 300301 300243 300220 300166 300176 300188 300269 300273 300144 300398 300229 300721 300722 300199 300200 300309 300632 300602 300601 300314 300177 300224 300669 300628 300291 300223 300456 300719 300718 300410 300547 300450 300649 300116 300171 300368 300145 300250 300288 300331 300338 300362 300407 300670 300693 300190 300230 300235 300297 300203 300465 300241 300280 300684 300686 300493 300324 300570 300569 300592 300657 300285 300358 300457 300460 300249 300533 300680 300681 300682 300462 300373 300729 300665 300148 300153 300154 300158 300178 300267 300591 300616 300406 300405 300672 300354 300385 300504 300505 300433 300432 300136 300711 300712 300501 300506 300117 300312 300195 300516 300402 300189 300514 300279 300126 300401 300537 300403 300464 300683 300664 300198 300491 300490 300420 300424 300327 300423 300427 300426 300579 300687 300688 300689 300699 300308 300307 300227 300228 300431 300430 300371 300238 300155 300329 300168 300245 300727 300726 300651 300194 300614 300613 300343 300240 300557 300483 300676 300675 300643 300674 300205 300207 300206 300315 300356 300409 300408 300691 300692 300283 300256 300135 300197 300626 300218 300306 300399 300179 300182 300123 300147 300523 300522 300563 300566 300598 300599 300561 300562 300500 300512 300530 300531 300548 300606 300573 300545 300574 300585 300733 300448 300449 300400 300404 300447 300627 300624 300630 300477 300473 300653 300701 300700 300161 300258 300697 300708 300478 300471 300581 300582 300603 300617 300611 300615 300622 300635 300192 300234 300489 300544 300658 300377 300453 300417 300421 300461 300469 300472 300527 300372 300470 300647 300272 300150 300479 300539 300219 300507 300508 300310 300482 300193 300535 300341 300475 300474 300351 300685 300724 300725 300183 300723 300646 300293 300551 300518 300259 300340 300639 300352 300355 300359 300620 300636 300652 300284 300286 300397 300129 300130 300133 300134 300695 300254 300261 300264 300263 300325 300553 300379 300587 300667 300540 300556 300208 300342 300345 300347 300580 300583 300734 300736 300619 300618 300575 300578 300434 300164 300167 300201 300232 300281 300282 300322 300323 300330 300175 300376 300378 300326 300594 300590 300253 300302 300172 300468 300476 300393 300392 300732 300731 300717 300715 300287 300416 300510 300631 300629 300743 300552 300294 300337 300660 300661 300300 300633 300640 300735 300209 300212 300211 300213 300214 300215 300525 300526 300648 300650 300554 300555 300517 300369 300160 300480 300454 300519 300509 300543 300455 300458 300387 300137 300139 300498 300702 300586 300210 300339 300346 300350 300741 300740 300705 300706 300728 300361 300360 300451 300452 300536 300673 300671 300296 300538 300248 300304 300511 300275 300366 300367 300446 300128 300217 300485 300170 300334 300415 300271 300313 300266 300231 300666 300196 300494 300703 300704 300481 300502 300542 300222 300138 300396 300173 300320 300612 300439 300442 300738 300549 300274 300181 300202 300236 300242 300428 300191 300233 300252 300644 300656 300336 300186 300382 300659 300696 300694 300375 300634 3006…
- UPC codes
- 15099590732103
Products covered by this recall
DxI 9000 Access Immunoassay Analyzer, Part Number C11137
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 15099590732103
Model or item number
On the rating label, the box or the instruction sheet.
- C11137 — C11137 UDI-DI
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Malfunction
Beckman Coulter, Inc. — This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
- Check whether the remedy is a software update, and if so confirm afterwards in the device menu that the version actually changed.
- If you are a clinic or facility, quarantine affected stock and record the correction in your recall log.
- Report device problems to the FDA through MedWatch — that database is where this recall came from.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Beckman Coulter, Inc. — 1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Beckman Coulter, Inc. medical devices recalled?
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected. Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 15099590732103; model number C11137 — C11137 UDI-DI. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE recall: Beckman Coulter has identified a software issue affecting DxC 500i Clinical
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a…
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE recall: Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of…
-
Class II · 2 months ago
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter DxC 500i Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23352, MicroScan LabPro V5.03 Panel Update 11 recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."