Beckman Coulter, Inc. CellMek SPS Sample Preparation System recall: Automated pipetting, diluting and specimen processing workstations for flow
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0412-2026
- Published
- November 5, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Beckman Coulter, Inc.
- Product type
- Medical devices
- Units affected
- 108 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- All serial numbers between BE43003 through BJ18014, manufactured from 12/03/2021 to 04/15/2025. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Products covered by this recall
CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Check your product
The notice names 3 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Model or item number
On the rating label, the box or the instruction sheet.
- C44603. Software versions 2.2.122.0
Serial number
On the rating plate, often next to the date of manufacture.
- between BE43003
- BJ18014
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Beckman Coulter, Inc. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Beckman Coulter, Inc. — 11800 Sw 147th Ave, Miami, FL, 33196-2500
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Beckman Coulter, Inc. medical devices recalled?
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment. Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: model number C44603. Software versions 2.2.122.0. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE recall: Beckman Coulter has identified a software issue affecting DxC 500i Clinical
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a…
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE recall: Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of…
-
Class II · 2 months ago
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter DxC 500i Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23352, MicroScan LabPro V5.03 Panel Update 11 recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."