Beckman Coulter, Inc. DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE recall: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
The hazard: Device malfunction
This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
What the notice says: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.
Symptoms and when to call a doctor
If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-0215-2026
- Published
- October 29, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Beckman Coulter, Inc.
- Product type
- Medical devices
- Units affected
- 160 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- C63522 UDI-DI: 14987666545089; C63521 UDI-DI: 14987666545072; C13252 UDI-DI: 15099590742331; Serial Numbers: 750198 750058 750222 750128 750138 750069 750050 750055 750179 750180 750181 750109 750168 750104 750211 750073 750210 750209 750178 750151 750153 750152 750155 750185 750187 750188 750195 750196 750201 750197 750199 750204 750202 750205 750083 750079 750186 750052 750230 750228 750140 750077 750135 750130 750167 750095 750105 750175 750132 750082 750113 750221 750219 750100 750099 750174 750053 750246 750134 750217 750216 750236 750233 750143 750081 750085 750071 750208 750207 750076 750131 750148 750223 750137 750116 750166 750241 750240 750183 750193 750200 750070 750087 750239 750107 750093 750124 750215 750141 750145 750144 750117 750108 750184 750173 750224 750218 750123 750129 750054 750206 750146 750101 750112 750194 750115 750157 750159 750160 750162 750125 750103 750229 750226 750234 750133 750048 750163 750182 750192 750189 750158 750250 750214 750110 750247 750154 750080 750251 750232 750231 750255 750238 750253 750074 750047 750102 750147 750142 750092 750094 750164 750165 750225 750212 750064 750065 750177 750086 750088 750042 750043 750254 750044 750056 750169 750045 750057 750127 750156
- UPC codes
- 14987666545089, 14987666545072, 15099590742331
Products covered by this recall
DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Module, DxC 500i (Part Number C13252)
Check your product
The notice names 6 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 14987666545089
- 14987666545072
- 15099590742331
Model or item number
On the rating label, the box or the instruction sheet.
- C63522
- C63521
- C13252
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Malfunction
Beckman Coulter, Inc. — This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
- If the device is implanted, or you depend on it daily, speak to your clinician before changing anything about how you use it.
- Check whether the remedy is a software update, and if so confirm afterwards in the device menu that the version actually changed.
- If you are a clinic or facility, quarantine affected stock and record the correction in your recall log.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Beckman Coulter, Inc. — 1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Beckman Coulter, Inc. medical devices recalled?
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected. Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 14987666545089, 14987666545072, 15099590742331; model number C63522, C63521, C13252. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE recall: Beckman Coulter has identified a software issue affecting DxC 500i Clinical
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a…
-
Class II · 11 days ago
Beckman Coulter Inc. DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE recall: Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of…
-
Class II · 2 months ago
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter DxC 500i Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously…
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
-
Class II · 2 months ago
Beckman Coulter Inc. REF # D23352, MicroScan LabPro V5.03 Panel Update 11 recall: Analyzers contain incorrect algorithms which may affect antimicrobial
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."