Class II U.S. Food and Drug Administration

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software recall: a software defect in which the Parameter Package (PP) import logic overwrites

Recall published 2 months ago

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

The hazard: Device malfunction

A malfunction was identified in this device. What matters most is whether your particular unit falls inside the affected model, lot or serial range.

What the notice says: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.

Symptoms and when to call a doctor

If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-2974-2026
Published
August 26, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Beckman Coulter Inc.
Product type
Medical devices
Distribution
Nationwide
Made in
United States
Lot / code info
1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
UPC codes
14987666545058, 14987666545065

Products covered by this recall

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Check your product

The notice names 7 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 14987666545058
  • 14987666545065

Model or item number

On the rating label, the box or the instruction sheet.

  • C63519
  • C63520
  • UDI
  • DI 14987666545058 2
  • DI 14987666545065

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Malfunction

Beckman Coulter Inc. — A malfunction was identified in this device. What matters most is whether your particular unit falls inside the affected model, lot or serial range.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
  2. If you are a clinic or facility, quarantine affected stock and record the correction in your recall log.
  3. Report device problems to the FDA through MedWatch — that database is where this recall came from.
  4. If the device is implanted, or you depend on it daily, speak to your clinician before changing anything about how you use it.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: Beckman Coulter Inc. — 250 S Kraemer Blvd, Brea, CA, 92821-6232

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Beckman Coulter Inc. medical devices recalled?

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 14987666545058, 14987666545065; model number C63519, C63520, UDI. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.