Recalls affecting District of Columbia
Notices that name District of Columbia in their distribution area, plus every nationwide recall — nationwide recalls reach District of Columbia too.
-
Class II · 10 months ago
Zimmer, Inc. Brand Name: Affixus Antegrade Femoral Nailing System Product Name recall: Affected implants may have undersized distal diameter along length
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk…
-
Class II · 10 months ago
Zimmer, Inc. Brand Name: Affixus Antegrade Femoral Nailing System Product Name recall: Affected implants may have undersized distal diameter along length
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk…
-
Class II · 10 months ago
Medtronic Neuromodulation Restore Clinician Programmer Application recall: Software issue where a Device Reset message displayed on the app was unable
Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume…
-
Class II · 10 months ago
CareFusion 303, Inc. BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer recall: Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to…
-
Class II · 10 months ago
Merck Sharp & Dohme LLC Noxafil Powdermix for Delayed-Release Oral Suspension, 300 mg recall: Presence of particulate matter: potential presence of metal particulates
Presence of particulate matter: potential presence of metal particulates in the product.
-
Class II · 10 months ago
Merck Emend for oral suspension, 125 mg, Single-Dose Kit recall: Presence of particulate matter: potential presence of metal particulates
Presence of particulate matter: potential presence of metal particulates in the product.
-
Class I · 10 months ago
Carousel Candies Choceur Cookie Butter Holiday Bark. Standup Pouch Bag. 5 oz recall: Undeclared pecan
Undeclared pecan
-
Class I · 10 months ago
Carousel Candies Choceur Pecan Cranberry & Cinnamon Holiday Bark. Standup Pouch Bag recall: Undeclared wheat
Undeclared wheat
-
Class II · 10 months ago
Chips Ahoy Baked brownie bites recall: Product Safety – choking threats
Product Safety – choking threats
-
Class II · 10 months ago
Gold Star Distribution All FDA-regulated products held at facility including drugs recall: Potential Salmonella contamination, presence of rodent and avian
Potential Salmonella contamination, presence of rodent and avian contamination and insanitary conditions during the storage process.
-
Class I · 10 months ago
GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system Power recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power…
-
Class II · 10 months ago
Dsaart LLC Brand Name: AART Gluteal Implant Product Name: Gluteal Implant recall: Due to manufactured products that deviated from established specifications
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket…
-
Class II · 10 months ago
Dsaart LLC Brand Name: AART Pectoral Implant Product Name: Pectoral Implant Model/ recall: Due to manufactured products that deviated from established specifications
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket…
-
Class II · 10 months ago
Medline Industries, LP Forcep, STD STR, Sterile Kelly, 5.5 recall: The items may have weak packaging seals that may result in a breach
The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if…
-
Class II · 10 months ago
Medline Industries, LP Scissors, OR, Sharp/Blunt, 5.5", Sterile recall: The items may have weak packaging seals that may result in a breach
The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if…
-
Class II · 10 months ago
Dsaart LLC Brand Name: AART Chin Implant Product Name: Chin Implant Model/Catalog recall: Due to manufactured products that deviated from established specifications
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket…
-
Class I · 10 months ago
GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power…
-
Class II · 10 months ago
Dsaart LLC Brand Name: AART Silicone Carving Implant Product Name: Silicone recall: Due to manufactured products that deviated from established specifications
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket…
-
Class III · 10 months ago
Datascope Corp. Cardiosave Hybrid recall: The IFU addendum updates the Vibration and Shock Table to reference
The IFU addendum updates the Vibration and Shock Table to reference the correct standards.
-
Class II · 10 months ago
Dsaart LLC Brand Name: AART Malar Implant Product Name: Malar Implant Model/ recall: Due to manufactured products that deviated from established specifications
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket…
-
Class II · 10 months ago
Dsaart LLC Brand Name: AART Calf Implant Product Name: Calf Implant Model/ recall: Due to manufactured products that deviated from established specifications
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket…
-
Class I · 10 months ago
GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power…
-
Class III · 10 months ago
Datascope Corp. Cardiosave Rescue recall: The IFU addendum updates the Vibration and Shock Table to reference
The IFU addendum updates the Vibration and Shock Table to reference the correct standards.
-
Class II · 10 months ago
Abbott Laboratories Alinity ci-series System Control Module , REF: 03R70-01, 03R70-20 recall: Abbott Laboratories is recalling their Alinity ci-series System Control
Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the…