Class I U.S. Food and Drug Administration

GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia

Recall published 10 months ago

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-0914-2026
Published
December 24, 2025
Risk level
Class I
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
GE Medical Systems China Co., Ltd.
Product type
Medical devices
Units affected
888 units
Distribution
Nationwide
Made in
China
Lot / code info
REF 1012-9750-000: GTIN 00840682145596; REF 1012-9750-002: GTIN 00840682146470; REF 1012-9755-000: GTIN 00840682146425; REF 1012-9755-002: GTIN 00840682146463.
UPC codes
00840682145596, 00840682146470, 00840682146425, 00840682146463

Products covered by this recall

GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000; Carestation 750 A2, REF 1012-9750-002; Carestation 750c A1, REF 1012-9755-000; Products for which affected PMB can be used as a spare part: Carestation 750c A2, REF 1012-9755-002

Check your product

The notice names 9 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00840682145596
  • 00840682146470
  • 00840682146425
  • 00840682146463

Model or item number

On the rating label, the box or the instruction sheet.

  • 1012-9750-000
  • 1012-9750-002
  • 1012-9755-000
  • 1012-9755-002 — REF 1012-9750-000
  • 1012-9755-002

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

GE Medical Systems China Co., Ltd. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. Report device problems to the FDA through MedWatch.
  3. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  4. Ask the supplier or clinic that provided it what the remedy is in your case.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: GE Medical Systems China Co., Ltd. — Dev. Zone, Wuxi

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was GE Medical Systems China Co., Ltd. medical devices recalled?

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00840682145596, 00840682146470, 00840682146425; model number 1012-9750-000, 1012-9750-002, 1012-9755-000. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.

Related recalls