GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0913-2026
- Published
- December 24, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- GE Medical Systems China Co., Ltd.
- Product type
- Medical devices
- Units affected
- 1277 units
- Distribution
- Nationwide
- Made in
- China
- Lot / code info
- REF 1012-9620-222: GTIN 00195278626301; REF 1012-9620-000: GTIN 00840682103985; REF 1012-9620-200: GTIN 00195278439536; REF 1012-9620-202: GTIN 00195278626158; REF 1012-9620-212: GTIN 00195278626561; REF 1012-9650-222: GTIN 00195278626592; REF 1012-9650-000: GTIN 00840682103947; REF 1012-9650-200: GTIN 00195278439529; REF 1012-9650-202: GTIN 00195278626585; REF 1012-9650-212: GTIN 00195278625687; REF 1012-9655-202: GTIN 00195278625953; REF 1012-9620-002: GTIN 00840682124546; REF 1012-9620-012: GTIN 00195278569677; REF 1012-9650-002: GTIN 00840682124560; REF 1012-9650-012: GTIN 00195278569684; REF 1012-9655-000: GTIN 00840682103954; REF 1012-9655-002: GTIN 00840682124539; REF 1012-9655-200: GTIN 00195278439543.
- UPC codes
- 00195278626301, 00840682103985, 00195278439536, 00195278626158, 00195278626561, 00195278626592, 00840682103947, 00195278439529, 00195278626585, 00195278625687, 00195278625953, 00840682124546, 00195278569677, 00840682124560, 00195278569684, 00840682103954, 00840682124539, 00195278439543
Products covered by this recall
GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 610 A1, REF 1012-9620-222; Carestation 620 A1 REF 1012-9620-000; Carestation 620 A1, REF 1012-9620-200; Carestation 620 A1, REF 1012-9620-202; Carestation 620 SE A1, REF 1012-9620-212; Carestation 630 A1, REF 1012-9650-222; Carestation 650 A1, REF 1012-9650-000; Carestation 650 A1, REF 1012-9650-200; Carestation 650 A1, REF 1012-9650-202; Carestation 650 SE A1, REF 1012-9650-212; Carestation 650c A1, REF 1012-9655-202; Anesthesia Gas Machine Products for which affected PMB can be used as a spare part: Carestation 620 A2, REF 1012-9620-002 00840682124546 Carestation 620 SE A2, REF 1012-9620-012 00195278569677 Carestation 650 A2, REF 1012-9650-002 00840682124560 Carestation 650 SE A2, REF 1012-9650-012 00195278569684 Carestation 650c A1, REF 1012-9655-000 00840682103954 Carestation 650c A2, REF 1012-9655-002 00840682124539 Carestation 650c A1, REF 1012-9655-200 00195278439543
Check your product
The notice names 37 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00195278626301
- 00840682103985
- 00195278439536
- 00195278626158
- 00195278626561
- 00195278626592
- 00840682103947
- 00195278439529
- 00195278626585
- 00195278625687
- 00195278625953
- 00840682124546
- 00195278569677
- 00840682124560
- 00195278569684
- 00840682103954
- 00840682124539
- 00195278439543
Model or item number
On the rating label, the box or the instruction sheet.
- 1012-9620-222
- 1012-9620-000
- 1012-9620-200
- 1012-9620-202
- 1012-9620-212
- 1012-9650-222
- 1012-9650-000
- 1012-9650-200
- 1012-9650-202
- 1012-9650-212
- 1012-9655-202
- RE
- 1012-9620-002
- 1012-9620-012
- 1012-9650-002
- 1012-9650-012
- 1012-9655-000
- 1012-9655-002
- 1012-9655-200
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
GE Medical Systems China Co., Ltd. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: GE Medical Systems China Co., Ltd. — Dev. Zone, Wuxi
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was GE Medical Systems China Co., Ltd. medical devices recalled?
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00195278626301, 00840682103985, 00195278439536; model number 1012-9620-222, 1012-9620-000, 1012-9620-200. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 5 months ago
GE Medical Systems China Co., Ltd. MAC 5 A4 Resting ECG Analysis System 855001-001 recall: Software in development inadvertently made available for customer use. Use
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being…
-
Class I · 10 months ago
GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power…
-
Class I · 10 months ago
GE Medical Systems China Co., Ltd. GE Healthcare Carestation anesthesia system Power recall: Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power…
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…