Class II U.S. Food and Drug Administration

Tobii Dynavox Llc TD Navio Maxi Speech Generating Device Model/ recall: Under certain conditions, the battery mounting tabs may become damaged

Recall published 2 months ago

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-2960-2026
Published
August 26, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Tobii Dynavox Llc
Product type
Medical devices
Distribution
Nationwide
Made in
United States
Lot / code info
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).

Products covered by this recall

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Check your product

The notice names 4 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

Lot or batch code

Usually ink-stamped near the seam, the cap or the bottom of the pack.

  • 20260615

Model or item number

On the rating label, the box or the instruction sheet.

  • ion batch 20260615

Serial number

On the rating plate, often next to the date of manufacture.

  • TDNL1-240904000016
  • TDNL1-26061707

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Tobii Dynavox Llc — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  3. Ask the supplier or clinic that provided it what the remedy is in your case.
  4. If you are a facility, quarantine affected stock and log the correction.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: Tobii Dynavox Llc — 2100 Wharton St Ste Ste 400, Pittsburgh, PA, 15203-1691

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Tobii Dynavox Llc medical devices recalled?

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: lot code 20260615; model number ion batch 20260615. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.