Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0088-2027
- Published
- October 14, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Medtronic MiniMed, Inc.
- Product type
- Medical devices
- Units affected
- 25
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770, FB00015162, FB00016429, FB00014760 were returned via complaint process or shipped to Marketing Team Members).
- UPC codes
- 00763000848156, 00763000848149
Products covered by this recall
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
Check your product
The notice names 11 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00763000848156
- 00763000848149
Model or item number
On the rating label, the box or the instruction sheet.
- MMT-8062
Serial number
On the rating plate, often next to the date of manufacture.
- FB00015839
- FB00016511
- FB00014455
- FB000
- FB00011770
- FB00015162
- FB00016429
- FB00017271
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Medtronic MiniMed, Inc. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Medtronic MiniMed, Inc. — 18000 Devonshire St, Northridge, CA, 91325-1219
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Medtronic MiniMed, Inc. medical devices recalled?
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00763000848156, 00763000848149; model number MMT-8062. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782 recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712 recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 3 months ago
Medtronic MiniMed, Inc. MiniMed Go App, Model: MMT-8600 , MMT-8601 recall: For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the…
-
Class II · 3 months ago
Medtronic MiniMed, Inc. MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886 recall: A software anomaly can occur when the pump is paired with the Instinct
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering…
-
Class II · 7 months ago
Medtronic MiniMed, Inc. MiniMed 720G Insulin Pump recall: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and…