Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0078-2027
- Published
- October 14, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems
- Product type
- Medical devices
- Units affected
- 539
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- 1. Model Number: 728333 (U.S.); UDI-DI: 1. 00884838101111, 2. 00884838127876; Serial Numbers: 12055, 12054, 12052, 398006, 12044, 12045, 10237, 10056, 10194, 12059, 10278, 10319, 10143, 10081, 12047, 12048, 10198, 12046, 12043, 10262, 12042, 12041, 10186, 10241, 12039, 10317, 10316, 10264, 10127, 12038, 12037, 396014, 12036, 398005, 10016, 12034, 10076, 10074, 12035, 10021, 10018, 10023, 12032, 12029, 398003, 398004, 396013, 12027, 12026, 12023, 12025, 12024, 12021, 12022, 12017, 12019, 12018, 12020, 10312, 10311, 10310, 10305, 12016, 10308, 10306, 10307, 12014, 12013, 10304, 10302, 12012, 12011, 12010, 396012, 12007, 10301, 12009, 10300, 10299, 12008, 12006, 12005, 10298, 12002, 12003, 10294, 10292, 12001, 12000, 10293, 10290, 396011, 396010, 10291, 10289, 10279, 10288, 10287, 10284, 10286, 10281, 10285, 10271, 10276, 396009, 10280, 10275, 10273, 10277, 396008, 396006, 396007, 10274, 10272, 10270, 10269, 10252, 10154, 10268, 10267, 10261, 10266, 10265, 10219, 10263, 10260, 10157, 10259, 10224, 10240, 10257, 10246, 10258, 10253, 10256, 10251, 10254, 10243, 10250, 10248, 10247, 10244, 10249, 10245, 10236, 10223, 10239, 10242, 10235, 10234, 10221, 10231, 10209, 10232, 10228, 10226, 10229, 10230, 10220, 10217, 10227, 10225, 10218, 10222, 10212, 10214, 10215, 10213, 10211, 10210, 396004, 10208, 10207, 10206, 10204, 10202, 10197, 10203, 10205, 10200, 10201, 10199, 10196, 10195, 10193, 10175, 10190, 10189, 10188, 10191, 10187, 10184, 10185, 10183, 10181, 10180, 10182, 10179, 10178, 10177, 10174, 10176, 10173, 10172, 10171, 10170, 10168, 10169, 10167, 10163, 10165, 10166, 10158, 10161, 10160, 10148, 10159, 10156, 10155, 10153, 10152, 10150, 10151, 10149, 10147, 10146, 10145, 10144, 10141, 10142, 10140, 10138, 10139, 10137, 10134, 10136, 10135, 10133, 10132, 10130, 10131, 10128, 10129, 10126, 10049, 10124, 10119, 10123, 10121, 10120, 10118, 10122, 10032, 10116, 10113, 10115, 10114, 10111, 10112, 10110, 10104, 396003, 396001, 10109, 10108, 10107, 10106, 10105, 10097, 10103, 10100, 10102, 10101, 10098, 10099, 10090, 10096, 10095, 10093, 10094, 10091, 10086, 10088, 10085, 10089, 10087, 10084, 10082, 10080, 10077, 10078, 10075, 10069, 10072, 10071, 10070, 10067, 10068, 10066, 10065, 10062, 10061, 10060, 10057, 10058, 10052, 10055, 10053, 10050, 10047, 10045, 10048, 10044, 10042, 10043, 10041, 10040, 10039, 10038, 10036, 10035, 10034, 10031, 10033, 10027, 10024, 10020, 10019, 10017, 10015, 10013, 10117, 10079, 10314, 12028, 12015, 10303, 10233, 10162, 12033, 12031, 12030, 10092, 10046, 10125, 10063, 10025; 2. Model Number: 728340 (Outside U.S.); UDI-DI: 00884838111103; Serial Numbers: 397035, 397004, 397168, 397167, 397164, 397159, 397160, 397155, 397153, 397152, 397150, 397151, 397148, 397147, 397146, 397145, 397141, 397142, 397139, 397140, 397137, 397133, 397132, 397130, 397126, 397125, 397123, 397121, 397120, 397117, 397116, 397115, 397112, 397110, 397108, 397107, 397106, 397102, 397101, 397099, 397098, 397096, 397095, 397093, 397091, 397085, 397089, 397088, 397087, 397086, 397084, 397083, 397082, 397080, 397081, 397079, 397077, 397076, 397075, 397073, 397072, 397071, 397070, 397067, 397066, 397065, 397062, 397064, 397063, 397060, 397061, 397059, 397058, 397056, 397055, 397054, 397053, 397050, 397049, 397047, 397048, 397046, 397045, 397043, 397044, 397041, 397040, 397039, 397038, 397037, 397036, 397034, 397033, 397032, 397026, 397031, 397029, 397028, 397027, 397025, 397024, 397023, 397022, 397021, 397020, 397018, 397019, 397016, 397017, 397015, 397014, 397012, 397009, 397013, 397011, 397003, 397010, 397008, 397006, 397002, 397005, 397007, 397105, 397030, 397128, 397119, 397097; 3. Model Number: 728344 (Outside U.S.); UDI-DI: 00884838117822; Serial Numbers: 397170, 397162, 397158, 397149, 397166, 397165, 397163, 397161, 397156, 397143, 397136, 397135, 397131, 397129, 397127, 397122, 397124, 397118, 397114, 397113, 397111, 397104, 397100, 397094, 397092, 397090, 397078, 397074, 397068, 397057, 397103; 4. Model Number: 728345 (Outside U.S…
- UPC codes
- 00884838101111, 00884838127876, 00884838111103, 00884838117822
Products covered by this recall
Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);
Check your product
The notice names 25 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838101111
- 00884838127876
- 00884838111103
- 00884838117822
Model or item number
On the rating label, the box or the instruction sheet.
- 1. 728333
- 728333
- 728340
- 728344
- 728345
Serial number
On the rating plate, often next to the date of manufacture.
- 12055
- 12054
- 12052
- 398006
- 12044
- 12045
- 397035
- 397004
- 397168
- 397167
- 397164
- 397170
- 397162
- 397158
- 397149
- 397166
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: PHILIPS MEDICAL SYSTEMS — 222 Jacobs St, Cambridge, MA, 02141-2289
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems medical devices recalled?
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion. Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838101111, 00884838127876, 00884838111103; model number 1. 728333, 728333, 728340. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 3 months ago
Philips Medical Systems IQon Spectral CT IQon Spectral CT Upgrades recall: The clinical CT scan may not automatically trigger after the Bolus Tracking
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking…
-
Class II · 3 months ago
Philips Medical Systems Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP recall: The clinical CT scan may not automatically trigger after the Bolus Tracking
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking…
-
Class II · 3 months ago
Philips Medical Systems Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 recall: The clinical CT scan may not automatically trigger after the Bolus Tracking
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking…
-
Class II · 6 months ago
Philips Spectral CT on Rails recall: Philips has identified three software issues: 1. During a continuous CT (CCT)
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the…
-
Class II · 12 months ago
Philips Medical Systems Model: Incisive CT for Brazil SKD recall: Issue 1: The potential for unintentional continued gantry/couch movement
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2…
-
Class II · 12 months ago
Philips Medical Systems Model: CT 5300 recall: Issue 1: The potential for unintentional continued gantry/couch movement
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2…