Therakos Development Limited Brand Name: Cellex Photopheresis System Product Name recall: Increased difficulty in installing the centrifuge bowl onto the centrifuge
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
The hazard: Sterility or particulate concern
Particulate matter or a sterility failure prompted this notice, the most common reason injectable and ophthalmic products are withdrawn.
What the notice says: Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
Why it matters: The product was made or packaged in conditions that cannot assure sterility, or visible particles were found in it. For anything injected, infused or used in the eye, that is a direct route for contamination into the body.
Symptoms and when to call a doctor
Watch for fever, chills, pain, redness, swelling or discharge at an injection site, and for a red, painful or suddenly blurred eye after using an eye product. Any of these needs same-day medical attention. Do not wait to see whether it settles, and tell the clinician which product and lot you used so it can be reported against the recall.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-1822-2026
- Published
- April 22, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Therakos Development Limited
- Product type
- Medical devices
- Units affected
- 11,499 units
- Distribution
- Nationwide
- Made in
- Ireland
- Lot / code info
- Model/Catalog Number: CLXUSA; UDI-DI: 20705030200003; Lot Numbers: N301, N339, N340, N341, N342, N343, N344, N345, N346, N347; P302, P303, P304, P305, P306, P307, P308, P309, P310, P311, P312, P313, P314, P319, P320.
- UPC codes
- 20705030200003
Products covered by this recall
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable
Check your product
The notice names 13 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 20705030200003
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- N301
- N339
- N340
- N341
- N342
- N343
- N344
- N345
- N346
- N347
Model or item number
On the rating label, the box or the instruction sheet.
- CLXUSA Software Version
- CLXUSA
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Sterility
Therakos Development Limited — Particulate matter or a sterility failure prompted this notice, the most common reason injectable and ophthalmic products are withdrawn.
- Stop using the affected lot and check the lot number on your own carton or vial against the codes listed above.
- Call your prescriber or pharmacy today about a replacement from an unaffected lot — do not simply skip doses of a medicine you depend on.
- If you are a clinic or pharmacy, quarantine the stock and work it through your usual recall log.
- Report any infection or reaction you think the product caused to FDA MedWatch.
What the company owes you
A recalled medicine is the manufacturer’s cost to carry, not yours or your pharmacist’s.
- The purchase price or co-pay is refundable, and so is the cost of returning the product if they ask you to ship it.
- Ask the pharmacy to handle the claim where they can — they process recall returns in bulk and it saves you the correspondence.
- Keep the carton, the leaflet and the pharmacy label together until the claim is closed.
- Costs beyond the price of the medicine need to go to the company directly, in writing, with the recall number quoted.
Company contact: THERAKOS DEVELOPMENT LIMITED — Suite 1 Plaza 211, Blanchardstown Corporate Park 2 Blanchardstown Road North, Blanchardstown
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Therakos Development Limited medical devices recalled?
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss. Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 20705030200003; lot code N301, N339, N340; model number CLXUSA Software Version, CLXUSA. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Stop using the affected lot and check the lot number on your own carton or vial against the codes listed above.
Can I get my money back?
The purchase price or co-pay is refundable, and so is the cost of returning the product if they ask you to ship it.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…
-
Class II · 11 days ago
Onkos Surgical Inc. My3D recall: Personalized pelvic implants inadvertently mixed up and distributed
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.