Ortho-Clinical Diagnostics, Inc. Vitros Chemistry Products LAC Slides recall: VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2560-2025
- Published
- September 24, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Ortho-Clinical Diagnostics, Inc.
- Product type
- Medical devices
- Units affected
- 2,320 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Catalog number 8150112; UDI# 10758750004454; Coating 0130 and above, this includes Lot numbers 3533-0130-2947, 3533-0130-2972, 3533-0130-3001, 3533-0130-3002, 3533-0130-3048, 3533-0130-3084, 3533-0131-5887, 3533-0131-5958, 3533-0131-5962, 3533-0131-6058 ; Expiration date range 01-Oct-2026 to 01-Jan-2027
- UPC codes
- 10758750004454
Products covered by this recall
VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
Check your product
The notice names 8 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 10758750004454
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 3533-0130-2947
- 3533-0130-2972
- 3533-0130-3001
- 3533-0130-3002
- 3533-01
Best-by / use-by date
Printed beside the lot code, not on the price sticker.
- 320
Model or item number
On the rating label, the box or the instruction sheet.
- 8150112
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Ortho-Clinical Diagnostics, Inc. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Ortho-Clinical Diagnostics, Inc. — 100 Indigo Creek Dr, Rochester, NY, 14626-5101
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Ortho-Clinical Diagnostics, Inc. medical devices recalled?
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 10758750004454; lot code 3533-0130-2947, 3533-0130-2972, 3533-0130-3001; best-by date 320; model number 8150112. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 1+ years ago
Ortho-Clinical Diagnostics, Inc. Chemistry Products LAC Slides recall: VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure…
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…