Medtronic O-arm O2 Imaging System recall: Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging
Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2451-2026
- Published
- June 24, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Medtronic Navigation, Inc.-Boxborough
- Product type
- Medical devices
- Units affected
- 589 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model Number: BI70002000. GTIN (Serial Numbers): 00643169639683 (C1640), 00643169651814 (C3815, C3816, C3817, C3818, C3891, C3898, C3899, C3900, C3901, C3902, C3903, C3905, C3990, C3997, C3998, C3999, C4001, C4002, C4003, C4004, C4059, C4060, C4061, C4062, C4063, C4064, C4065, C4066, C4067, C4068, C4116, C4117, C4118, C4119, C4120, C4121, C4122, C4123, C4130, C4201, C4202, C4203, C4356, C4357, C4360, C4361), 00763000132378 (C2096, C2187), 00763000355555 (C2768), 00763000481599 (C3895, C3996), 00763000496586 (C3794, C3896, C4023, C4031, C4032, C4043, C4088, C4092), 00763000496593 (C3887, C3889, C3978, C3988, C3991, C3994, C4041, C4109, C4110, C4111, C4161), 00763000496609 (C3802), 00763000496630 (C4035), 00763000496647 (C4012, C4126, C4129), 00763000496661 (C3824, C3987, C4071, C4072), 00763000496678 (C3879), 00763000496692 (C3799, C3823, C3897, C3979, C4027, C4037, C4044, C4125), 00763000496708 (C3825), 00763000496715 (C4074, C4128), 00763000496722 (C3812, C3822, C3828, C3881, C4034, C4036, C4038, C4039, C4160), 00763000616304 (C3795), 00763000616397 (C3800), 00763000616441 (C3811), 00763000616458 (C3783, C3784), 00763000616564 (C3316, C3735, C3777, C3778, C3779, C3780, C3781, C3786, C3787, C3790, C3791, C3792, C3793, C3796, C3797, C3798), 00763000709365 (C4307), 00763000709372 (C4221, C4284, C4294, C4301, C4354, C4370, C4399), 00763000709389 (C4166, C4295, C4402), 00763000709396 (C4394), 00763000709402 (C4149, C4150), 00763000709419 (C4395), 00763000709426 (C4115, C4248, C4304), 00763000709457 (C4293, C4298,C4403), 00763000709464 (C4368), 00763000709488 (C4204, C4205, C4302, C4363, C4366), 00763000709518 (C4148, C4167, C4222, C4223, C4224, C4225, C4285, C4292, C4303, C4309, C4362, C4369, C4374, C4375), 00763000709532 (C4206, C4208, C4209, C4236, C4245, C4283, C4286, C4287, C4288, C4289, C4290, C4308, C4310, C4372, C4373, C4377, C4378), 00763000709556 (C4390), 00763000709570 (C4249, C4250), 00763000709594 (C4396), 00763000709631 (C3883, C3884, C3885, C3907, C3908, C3909, C3910, C3911, C3912, C3913, C3914, C3915, C3916, C3918, C3919, C3920, C3921, C3922, C3923, C3924, C3925, C3926, C3927, C3928, C3929, C3930, C3931, C3932, C3933, C3934, C3935, C3936, C3937, C3938, C3939, C3940, C3941, C3942, C3943, C3944, C3945, C3946, C3947, C3948, C3949, C3950, C3951, C3953, C3954, C3955, C3956, C3957, C3958, C3959, C3960, C3961, C3962, C3963, C3964, C3965, C3966, C3967, C3968, C3969, C3970, C3971, C3972, C3973, C3974, C3975, C3976, C3977, C3982, C3983, C3984, C4014, C4015, C4016, C4017, C4018, C4019, C4020, C4021, C4022, C4028, C4045, C4046, C4047, C4048, C4049, C4050, C4051, C4052, C4053, C4075, C4076, C4077, C4078, C4079, C4080, C4081, C4082, C4083, C4084, C4085, C4086, C4087, C4089, C4090, C4093, C4094, C4095, C4096, C4097, C4098, C4100, C4101, C4102, C4103, C4104, C4105, C4106, C4107, C4108, C4131, C4132, C4133, C4134, C4135, C4136, C4137, C4138, C4139, C4140, C4141, C4142, C4143, C4144, C4145, C4146, C4151, C4152, C4153, C4154, C4155, C4156, C4157, C4158, C4162, C4163, C4168, C4169, C4170, C4171, C4172, C4173, C4174, C4175, C4176, C4177, C4178, C4179, C4180, C4181, C4182, C4183, C4184, C4185, C4186, C4187, C4188, C4190, C4191, C4193, C4194, C4195, C4196, C4197, C4198, C4199, C4200, C4210, C4211, C4212, C4213, C4214, C4215, C4217, C4218, C4219, C4220, C4226, C4227, C4228, C4229, C4230, C4231, C4232, C4233, C4234, C4235, C4237, C4238, C4239, C4240, C4241, C4242, C4243, C4244, C4246, C4247, C4251, C4252, C4253, C4254, C4256, C4257, C4258, C4259, C4260, C4261, C4263, C4264, C4265, C4266, C4267, C4268, C4269, C4270, C4271, C4272, C4273, C4274, C4278, C4279, C4281, C4311, C4312, C4313, C4314, C4315, C4316, C4318, C4319, C4320, C4321, C4322, C4323, C4324, C4325, C4326, C4328, C4329, C4330, C4331, C4332, C4333, C4334, C4335, C4336, C4337, C4339, C4340, C4341, C4342, C4343, C4344, C4345, C4346, C4347, C4348, C4349, C4350, C4351, C4382, C4383, C4384, C4385, C4386, C4388, C4405, C4406, C4407, C4414, C4415, C4416, C4418, C4421, C4422, C4280), 0076…
- UPC codes
- 00643169639683, 00643169651814, 00763000132378, 00763000355555, 00763000481599, 00763000496586, 00763000496593, 00763000496609, 00763000496630, 00763000496647, 00763000496661, 00763000496678, 00763000496692, 00763000496708, 00763000496715, 00763000496722, 00763000616304, 00763000616397, 00763000616441, 00763000616458
Products covered by this recall
Medtronic O-arm O2 Imaging System. Model Number: BI70002000.
Check your product
The notice names 20 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00643169639683
- 00643169651814
- 00763000132378
- 00763000355555
- 00763000481599
- 00763000496586
- 00763000496593
- 00763000496609
- 00763000496630
- 00763000496647
- 00763000496661
- 00763000496678
- 00763000496692
- 00763000496708
- 00763000496715
- 00763000496722
- 00763000616304
- 00763000616397
- 00763000616441
- 00763000616458
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Medtronic Navigation, Inc.-Boxborough — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Medtronic Navigation, Inc.-Boxborough — 200 Beaver Brook Road, Boxborough, MA, 01719
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Medtronic Navigation, Inc.-Boxborough medical devices recalled?
Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware. Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00643169639683, 00643169651814, 00763000132378. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…
-
Class II · 11 days ago
Onkos Surgical Inc. My3D recall: Personalized pelvic implants inadvertently mixed up and distributed
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.