Medtronic MiniMed, Inc. Paradigm REAL-Time Revel Insulin Pump recall: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1750-2026
- Published
- April 8, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Medtronic MiniMed, Inc.
- Product type
- Medical devices
- Units affected
- 26,647
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Revel Insulin Pump: 00613994951502, 00763000161262, 00613994743930, 00643169077737, 00613994904485, 00643169202399, 00643169151369, 00643169478602, 00643169513914, 00643169933170, 00763000161132, 00643169464520: Artwork, CRB D Process, Paradigm REAL-Time Revel Insulin Pump UG (MP6025306-011A 1); Artwork, CRB D Process, Instr, X23 revel, EN (MP6025306-019 1)
- UPC codes
- 00613994951502, 00763000161262, 00613994743930, 00643169077737, 00613994904485, 00643169202399, 00643169151369, 00643169478602, 00643169513914, 00643169933170, 00763000161132, 00643169464520
Products covered by this recall
Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
Check your product
The notice names 12 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00613994951502
- 00763000161262
- 00613994743930
- 00643169077737
- 00613994904485
- 00643169202399
- 00643169151369
- 00643169478602
- 00643169513914
- 00643169933170
- 00763000161132
- 00643169464520
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Medtronic MiniMed, Inc. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Medtronic MiniMed, Inc. — 18000 Devonshire St, Northridge, CA, 91325-1219
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Medtronic MiniMed, Inc. medical devices recalled?
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00613994951502, 00763000161262, 00613994743930. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782 recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712 recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 3 months ago
Medtronic MiniMed, Inc. MiniMed Go App, Model: MMT-8600 , MMT-8601 recall: For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the…
-
Class II · 3 months ago
Medtronic MiniMed, Inc. MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886 recall: A software anomaly can occur when the pump is paired with the Instinct
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering…