Medtronic MiniMed, Inc. MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886 recall: A software anomaly can occur when the pump is paired with the Instinct
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
The hazard: Device malfunction
This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
What the notice says: A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.
Symptoms and when to call a doctor
If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-2698-2026
- Published
- July 15, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Medtronic MiniMed, Inc.
- Product type
- Medical devices
- Units affected
- 778
- Made in
- United States
- Lot / code info
- UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
- UPC codes
- 7630006392701, 00763000521554, 00763000639273, 7630006392301, 00199150035687, 7630007325603, 7630007325601, 00763000639235, 7630007325604, 00763000384289, 00763000521530, 7630007325401, 00763000521547, 7630006392601, 7630006392201, 00763000639228, 00199150035663, 00763000639266, 7630007325404, 00199150035694
Products covered by this recall
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Check your product
The notice names 25 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 7630006392701
- 00763000521554
- 00763000639273
- 7630006392301
- 00199150035687
- 7630007325603
- 7630007325601
- 00763000639235
- 7630007325604
- 00763000384289
- 00763000521530
- 7630007325401
- 00763000521547
- 7630006392601
- 7630006392201
- 00763000639228
- 00199150035663
- 00763000639266
- 7630007325404
- 00199150035694
Model or item number
On the rating label, the box or the instruction sheet.
- MMT-1885
- MMT-1886
- MMT-1895
- MMT-1896 — UDI-DI
- GTI
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Malfunction
Medtronic MiniMed, Inc. — This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
- Report device problems to the FDA through MedWatch — that database is where this recall came from.
- If the device is implanted, or you depend on it daily, speak to your clinician before changing anything about how you use it.
- Check whether the remedy is a software update, and if so confirm afterwards in the device menu that the version actually changed.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Medtronic MiniMed, Inc. — 18000 Devonshire St, Northridge, CA, 91325-1219
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Medtronic MiniMed, Inc. medical devices recalled?
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA). A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 7630006392701, 00763000521554, 00763000639273; model number MMT-1885, MMT-1886, MMT-1895. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Related recalls
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782 recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 2 months ago
Medtronic MiniMed, Inc. MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712 recall: Loaner insulin pumps may have been cleaned with chemically incompatible
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress…
-
Class II · 3 months ago
Medtronic MiniMed, Inc. MiniMed Go App, Model: MMT-8600 , MMT-8601 recall: For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the…
-
Class II · 7 months ago
Medtronic MiniMed, Inc. MiniMed 720G Insulin Pump recall: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and…