Class II U.S. Food and Drug Administration

Medtronic MiniMed, Inc. InPen App, CFN: MMT-8060 and MMT-8061 recall: Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users

Recall published 1+ years ago

Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

The hazard: Device safety concern

A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

What the notice says: Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-2496-2025
Published
September 10, 2025
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Medtronic MiniMed, Inc.
Product type
Medical devices
Units affected
6816
Distribution
Georgia
Made in
United States
Lot / code info
UDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0

Products covered by this recall

InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)

What to do about Device

Medtronic MiniMed, Inc. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. Report device problems to the FDA through MedWatch.
  3. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  4. Ask the supplier or clinic that provided it what the remedy is in your case.

Claiming on a recalled device

Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.

  • Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
  • Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
  • If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
  • Log the date, the name and the reference number of every call. Device recalls can run for months.

Company contact: Medtronic MiniMed, Inc. — 18000 Devonshire St, Northridge, CA, 91325-1219

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Medtronic MiniMed, Inc. medical devices recalled?

Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been…

How do I know if I have the recalled product?

The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Device enforcement record, and contact the company if you are unsure.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.

Which states were affected?

The notice lists Georgia.