Class II U.S. Food and Drug Administration

Getinge 88-Series Washer- Disinfector recall: Potential for device to overheat during operation if the circulation pressure

Recall published 1+ years ago

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

The hazard: Fire or burn hazard

This product can overheat or catch fire, which is why the instruction is to unplug and stop using it rather than to keep an eye on it.

What the notice says: Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Why it matters: The product can overheat, ignite or cause a burn. Where lithium batteries are involved the failure is fast and very hot, and water does not put it out quickly.

Symptoms and when to call a doctor

Treat a burn by cooling it under running water for about 20 minutes — not with ice, butter or ointment — and cover it loosely with a clean non-fluffy cloth. Get medical help for burns that are larger than the person’s palm, that are deep or white or charred, that involve the face, hands, feet or genitals, or any burn to a baby or young child. If someone has inhaled smoke and is coughing, hoarse or short of breath, call 911.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-0016-2026
Published
October 8, 2025
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Getinge Disinfection Ab
Product type
Medical devices
Units affected
902 units
Distribution
Nationwide
Made in
Sweden
Lot / code info
Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.
UPC codes
07340153700109

Products covered by this recall

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Check your product

The notice names 1 code you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 07340153700109

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Fire

Getinge Disinfection Ab — This product can overheat or catch fire, which is why the instruction is to unplug and stop using it rather than to keep an eye on it.

  1. Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.
  2. Move it off anything that burns — a bed, a sofa, a carpet, a pile of paper — and away from exits.
  3. Do not leave it charging overnight or while the house is empty while you wait for the remedy.
  4. If it has already overheated, scorched or swelled, do not keep using it and do not try to repair it.

Claiming your refund, repair or replacement

The recalling company decides the remedy and funds it. You should never pay for a recall repair or for return shipping.

  • Use the company contact on this page first. Most firms run a dedicated recall page with a claim form that is far quicker than their main support line.
  • If the company stops responding, escalate to the CPSC. Firms under a recall order have reporting duties to the agency.
  • Have the model number, the date code and your photos ready. Many companies issue store credit by default — ask for cash or a card reversal if you would rather have it.
  • No receipt? Say so. Recall refunds are routinely paid without proof of purchase, because the hazard is what triggers the obligation.

Company contact: Getinge Disinfection Ab — Ljungadalsgatan 11, Vaxjo

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Getinge Disinfection Ab medical devices recalled?

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered. Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 07340153700109. If the code on your package is not on that list, your unit is outside the recall.