GE Medical Systems SCS GE Healthcare AW Server 3.2 ext.6.5 recall: GE HealthCare has become aware of a context synchronization issue in AW
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2553-2026
- Published
- July 1, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- GE Medical Systems SCS
- Product type
- Medical devices
- Units affected
- 340 servers
- Distribution
- Nationwide
- Made in
- France
- Lot / code info
- REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
- UPC codes
- 00840682102384
Products covered by this recall
GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
Check your product
The notice names 1 code you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00840682102384
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
GE Medical Systems SCS — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: GE Medical Systems SCS — 283 rue de la Miniere, Buc
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was GE Medical Systems SCS medical devices recalled?
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00840682102384. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 1+ years ago
GE Medical Systems, SCS AW Server 2.0, AW Server 3.0, AW Server 3.1, AW Server 3.2 recall: Firm has identified a security vulnerability in AW Server products
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity…
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…