FUJIFILM Healthcare Americas Corporation FDR Visionary Suite recall: It was found that the contacts of the electromagnetic contactors used inside
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
The hazard: Fire or burn hazard
A fire or burn risk is what prompted this recall. Most of these incidents happen while the product is charging or left running unattended.
What the notice says: It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Why it matters: The product can overheat, ignite or cause a burn. Where lithium batteries are involved the failure is fast and very hot, and water does not put it out quickly.
Symptoms and when to call a doctor
Treat a burn by cooling it under running water for about 20 minutes — not with ice, butter or ointment — and cover it loosely with a clean non-fluffy cloth. Get medical help for burns that are larger than the person’s palm, that are deep or white or charred, that involve the face, hands, feet or genitals, or any burn to a baby or young child. If someone has inhaled smoke and is coughing, hoarse or short of breath, call 911.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-1299-2026
- Published
- February 11, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- FUJIFILM Healthcare Americas Corporation
- Product type
- Medical devices
- Units affected
- 16 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
- UPC codes
- 04540217052226, 04540217057436, 04540217057450
Products covered by this recall
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Check your product
The notice names 11 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 04540217052226
- 04540217057436
- 04540217057450
Model or item number
On the rating label, the box or the instruction sheet.
- CH-200
Serial number
On the rating plate, often next to the date of manufacture.
- MP95A9F6A001
- MP95A9F6B001
- MP95A9F74001
- MP95AA783001
- MP95AA983001
- MP95AA983002
- MP95AA983003
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Fire
FUJIFILM Healthcare Americas Corporation — A fire or burn risk is what prompted this recall. Most of these incidents happen while the product is charging or left running unattended.
- Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.
- Move it off anything that burns — a bed, a sofa, a carpet, a pile of paper — and away from exits.
- Do not leave it charging overnight or while the house is empty while you wait for the remedy.
- If it has already overheated, scorched or swelled, do not keep using it and do not try to repair it.
How the remedy works
CPSC recalls specify one of three remedies, and which one you get is not negotiable — but how smoothly it goes often is.
- A refund means the purchase price back, sometimes prorated by age for durable goods. Ask how they calculated it if the figure looks low.
- Return shipping is the company’s cost. If you are asked to pay it, push back.
- A repair kit is shipped free and is designed to be fitted at home; if you cannot fit it, ask whether they will arrange installation.
- A replacement is usually the current model, not the recalled one. Confirm the replacement is not itself under recall.
Company contact: FUJIFILM Healthcare Americas Corporation — 81 Hartwell Ave Ste 300, Lexington, MA, 02421-3160
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was FUJIFILM Healthcare Americas Corporation medical devices recalled?
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke. It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 04540217052226, 04540217057436, 04540217057450; model number CH-200. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.
Can I get my money back?
A refund means the purchase price back, sometimes prorated by age for durable goods. Ask how they calculated it if the figure looks low.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · last month
Fujifilm Synapse Pacs software v7.4.310 recall: Software issue may cause the ruler measurement to be inaccurate
Software issue may cause the ruler measurement to be inaccurate.
-
Class II · 1+ years ago
FUJIFILM Healthcare Americas Corporation FDR Visionary Suite recall: It was found that the shaft for mounting the X-ray tube unit on the support
It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not…
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…