Erbeflo CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & recall: Devices providing sterile water/CO2/air to endoscopes were manufactured
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
The hazard: Sterility or particulate concern
This recall is about sterility assurance. It does not mean every unit is contaminated — it means the manufacturer cannot demonstrate that none of them is.
What the notice says: Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Why it matters: The product was made or packaged in conditions that cannot assure sterility, or visible particles were found in it. For anything injected, infused or used in the eye, that is a direct route for contamination into the body.
Symptoms and when to call a doctor
Watch for fever, chills, pain, redness, swelling or discharge at an injection site, and for a red, painful or suddenly blurred eye after using an eye product. Any of these needs same-day medical attention. Do not wait to see whether it settles, and tell the clinician which product and lot you used so it can be reported against the recall.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-2062-2026
- Published
- May 13, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Erbe Medical, LLC
- Product type
- Medical devices
- Units affected
- 260,962
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249; 20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243; 20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367; 20325-207/04065655000689/W4469939, W4469940, W4469941; 20325-248/04065655001006/W4467050, W4467051, W4467055, W4467056, W4467682, WO470898, WO470899, WO470900, WO470901, WO470902, WO470903, WO471252, WO471253, WO471254, WO471255, WO471256; 20325-208/04065655001099/W4467040, W4468271, WO468760
- UPC codes
- 04065655000375, 04065655000405, 04065655000436, 04065655000689, 04065655001006, 04065655001099
Products covered by this recall
ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208
Check your product
The notice names 20 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 04065655000375
- 04065655000405
- 04065655000436
- 04065655000689
- 04065655001006
- 04065655001099
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 20325-239
- 04065655000375
- W4464506
- W4464507
- W4464512
- W4464514
- W44645
Model or item number
On the rating label, the box or the instruction sheet.
- 20325-239
- 20325-240
- 20325-206
- 20325-207
- 20325-248
- 20325-208 — REF
- UDI-DI
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Sterility
Erbe Medical, LLC — This recall is about sterility assurance. It does not mean every unit is contaminated — it means the manufacturer cannot demonstrate that none of them is.
- Stop using the affected lot and check the lot number on your own carton or vial against the codes listed above.
- Do not use a vial or syringe with visible particles, cloudiness or discolouration, whether or not it is on the recall list.
- Call your prescriber or pharmacy today about a replacement from an unaffected lot — do not simply skip doses of a medicine you depend on.
- If you are a clinic or pharmacy, quarantine the stock and work it through your usual recall log.
Refunds and replacements
With medicines, a replacement supply matters more than cash, and that is what pharmacies are set up to provide.
- Ask the dispensing pharmacy for a like-for-like replacement from an unaffected lot. This is the fastest route and usually costs nothing.
- For a mail-order or online pharmacy, use the recall line listed on this page rather than the general support queue.
- If the recall voided a fill you had already paid for, your insurer will normally authorise a replacement ahead of schedule. Quote the recall number.
- If you paid out of pocket, ask for a refund of the co-pay or purchase price and keep the receipt the pharmacy issues.
Company contact: Erbe Medical, LLC — 405 W Geneva Dr, Tempe, AZ, 85282-2003
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Erbe Medical, LLC medical devices recalled?
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure. Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 04065655000375, 04065655000405, 04065655000436; lot code 20325-239, 04065655000375, W4464506; model number 20325-239, 20325-240, 20325-206. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Stop using the affected lot and check the lot number on your own carton or vial against the codes listed above.
Can I get my money back?
Ask the dispensing pharmacy for a like-for-like replacement from an unaffected lot. This is the fastest route and usually costs nothing.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 6 months ago
Erbeflo CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201 recall: Devices providing sterile water/CO2/air to endoscopes were manufactured
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more…
-
Class II · 6 months ago
Erbeflo 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222 recall: Devices providing sterile water/CO2/air to endoscopes were manufactured
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more…
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…