Endomagnetics Ltd Magseed Pro 12 cm soft tissue marker recall: Potential for contamination with cotton fibers
Potential for contamination with cotton fibers.
The hazard: High-powered magnet hazard
This recall concerns small, strong magnets that can detach. Teenagers are at risk as well as toddlers, because magnets get used to mimic tongue and lip piercings.
What the notice says: Potential for contamination with cotton fibers.
Why it matters: Loose high-powered magnets are dangerous when more than one is swallowed. They attract each other through the walls of the intestine and pinch the tissue trapped between them, an injury that usually needs surgery.
Symptoms and when to call a doctor
If you think a child has swallowed magnets — even if they seem entirely well — go to an emergency department and say magnets are involved. Early symptoms are indistinguishable from a stomach bug: stomach pain, vomiting, poor appetite. A single magnet may pass on its own; two or more rarely do, and the team needs to know how many are unaccounted for.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-0317-2026
- Published
- November 5, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Endomagnetics Ltd
- Product type
- Medical devices
- Units affected
- 540 units
- Distribution
- Nationwide
- Made in
- United Kingdom
- Lot / code info
- Model No. MS2-17-1-12. GTIN:15060391210466. UDI: 15060391210466250327-01, 15060391210466250415-07, 15060391210466250523-10, 15060391210466250612-50. Batch Number: 250327-01, 250415-07, 250523-10, 250612-50.
- UPC codes
- 15060391210466
Products covered by this recall
Magseed Pro 12 cm soft tissue marker. Model Number: MS2-17-1-12. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
Check your product
The notice names 5 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 15060391210466
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 250327-01
- 250415-07
- 250523-10
- 250612-50. — 540 units
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Magnets
Endomagnetics Ltd — This recall concerns small, strong magnets that can detach. Teenagers are at risk as well as toddlers, because magnets get used to mimic tongue and lip piercings.
- Put the product out of reach and count the magnets against the number the instructions say it should have.
- If any are unaccounted for and a child could have reached them, go to an emergency department and tell them magnets may have been swallowed.
- Make sure older children know that using magnets as fake piercings is a recognised route to a serious injury.
- Follow the remedy described on this page rather than keeping the item with the magnets glued back in.
How the remedy works
CPSC recalls specify one of three remedies, and which one you get is not negotiable — but how smoothly it goes often is.
- A refund means the purchase price back, sometimes prorated by age for durable goods. Ask how they calculated it if the figure looks low.
- Return shipping is the company’s cost. If you are asked to pay it, push back.
- A repair kit is shipped free and is designed to be fitted at home; if you cannot fit it, ask whether they will arrange installation.
- A replacement is usually the current model, not the recalled one. Confirm the replacement is not itself under recall.
Company contact: ENDOMAGNETICS LTD — St. Johns Innovation Park, Cambridge
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Endomagnetics Ltd medical devices recalled?
Potential for contamination with cotton fibers. Potential for contamination with cotton fibers.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 15060391210466; lot code 250327-01, 250415-07, 250523-10. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Put the product out of reach and count the magnets against the number the instructions say it should have.
Can I get my money back?
A refund means the purchase price back, sometimes prorated by age for durable goods. Ask how they calculated it if the figure looks low.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 12 months ago
Endomagnetics Ltd Magseed Pro 7 cm soft tissue marker recall: Potential for contamination with cotton fibers
Potential for contamination with cotton fibers.
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…