Cerenovus Inc Cerepak Detachable Coil System - Name/REF: Uniform 2MM X 2.5 recall: Detachable coils system may fail to detach, which could result in hemorrhagic
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0434-2026
- Published
- November 19, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Cerenovus Inc
- Product type
- Medical devices
- Units affected
- 12004
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- All devices/lots manufactured prior to and including August 2025 are affected. REF/UDI-DI: FCX100202/10886704083986, FCX100253/10886704083979, FCX100305/10886704083962, FCX100356/10886704083993, FCX100407/10886704084013, FCX100412/10886704084006, FCX100509/10886704084037, FCX100517/10886704084020, FCX100611/10886704084044, FCX100626/10886704084051, FCX100713/10886704084068, FCX100730/10886704084075, FCX100816/10886704084082, FCX100829/10886704084099, FCX140225/10886704084167, FCX140303/10886704084174, FCX140304/10886704084181, FCX140306/10886704084198, FCX140406/10886704084211, FCX140408/10886704084228, FCX140445/10886704084204, FCX140507/10886704084242, FCX140512/10886704084235, FCX140609/10886704084266, FCX140615/10886704084259, FCX140711/10886704084273, FCX140717/10886704084280, FCX140812/10886704084297, FCX140820/10886704084303, FCX140914/10886704084310, FCX140922/10886704084327, FCX141016/10886704084105, FCX141025/10886704084112, FCX141118/10886704084129, FCX141127/10886704084136, FCX141219/10886704084143, FCX141230/10886704084150, FCX180510/10886704084464, FCX180612/10886704084471, FCX180714/10886704084488, FCX180813/10886704084495, FCX180830/10886704084501, FCX180915/10886704084518, FCX180933/10886704084525, FCX181017/10886704084334, FCX181034/10886704084341, FCX181118/10886704084358, FCX181137/10886704084365, FCX181220/10886704084372, FCX181240/10886704084389, FCX181343/10886704084396, FCX181447/10886704084402, FCX181550/10886704084419, FCX181647/10886704084426, FCX181750/10886704084433, FCX181846/10886704084440, FCX181950/10886704084457, FCX182050/10886704082880, MCR091010/10886704082897, MCR091015/10886704082903, MCR091020/10886704082910, MCR091025/10886704082927, MCR091030/10886704082934, MCR091040/10886704082941, MCR091520/10886704082958, MCR091525/10886704082965, MCR091530/10886704082972, MCR091540/10886704082989, MCR092020/10886704084778, MCR092030/10886704082996, MCR092040/10886704083009, MCR092060/10886704083016, MCR092535/10886704083023, MCR092545/10886704083030, MCR092555/10886704083047, MCR093040/10886704083054, MCR093060/10886704083061, MCR093080/10886704083078, SCR120202/10886704083085, SCR120306/10886704083092, SCR120308/10886704083108, SCR120407/10886704083115, SCR120410/10886704083122, SCR120412/10886704083139, SCR120510/10886704083146, SCR120515/10886704083153, SCR120610/10886704083160, SCR120615/10886704083177, SCR120620/10886704083184, SCR120715/10886704083191, SCR120721/10886704083207, SCR120815/10886704083214, SCR120824/10886704083221, SCR120925/10886704083238, SCR121030/10886704083245, SCR121230/10886704083252, SCR122505/10886704083269, SCR122535/10886704083276, SCR123509/10886704083283, SCR123575/10886704083290, SHD100154/10886704083306, SHD100208/10886704083313, SHD100250/10886704083320, SHD100258/10886704083337, SHD100308/10886704083344, SHD100310/10886704083351, SHD100408/10886704083368, SHD100410/10886704083375, SHD100510/10886704083382, SHD100515/10886704083399, SHD100612/10886704083405, SHD100616/10886704083412, SHD100716/10886704083429, SHD100720/10886704083436, SHD100820/10886704083443, SHD100925/10886704083450, SHD180312/10886704083467, SHD180415/10886704083474, SHD180520/10886704083481, SHD180625/10886704083498, SHD180733/10886704083504, SHD180835/10886704083511, SHD180935/10886704083528, SHD181040/10886704083535, SHD181242/10886704083542, SHD181445/10886704083559, SHD181650/10886704083566, SHD181855/10886704083573, SHD182060/10886704083580, SHD182260/10886704083597, SHD182460/10886704083603, XCR120201/10886704083610, XCR120202/10886704083627, XCR120304/10886704083634, XCR120306/10886704083641, XCR120308/10886704083658, XCR120406/10886704083665, XCR120408/10886704083672, XCR120410/10886704083689, XCR120510/10886704083696, XCR120515/10886704083702, XCR120610/10886704083719, XCR120620/10886704083726, XCR122505/10886704083733, XCR122525/10886704083740, XCR122535/10886704083757, XCR123505/10886704083764, XCR123509/10886704083771, XCR123575/10886704083788, XHD100151/10886704083795, XHD10015…
- UPC codes
- 10886704083986, 10886704083979, 10886704083962, 10886704083993, 10886704084013, 10886704084006, 10886704084037, 10886704084020, 10886704084044, 10886704084051, 10886704084068, 10886704084075, 10886704084082, 10886704084099, 10886704084167, 10886704084174, 10886704084181, 10886704084198, 10886704084211, 10886704084228
Products covered by this recall
CEREPAK Detachable Coil System - Name/REF: UNIFORM 2MM X 2.5 CM/FCX100202, UNIFORM 2.5MM X 3.3 CM/FCX100253, UNIFORM 3MM X 5.4 CM/FCX100305, UNIFORM 3.5MM X 6.6 CM/FCX100356, UNIFORM 4MM X 7.5 CM/FCX100407, UNIFORM 4MM X 11.5 CM/FCX100412, UNIFORM 5MM X 9.7 CM/FCX100509, UNIFORM 5MM X 17 CM/FCX100517, UNIFORM 6MM X 11.9 CM/FCX100611, UNIFORM 6MM X 26 CM/FCX100626, UNIFORM 7MM X 13.9 CM/FCX100713, UNIFORM 7MM X 30 CM/FCX100730, UNIFORM 8MM X 16.1 CM/FCX100816, UNIFORM 8MM X 29 CM/FCX100829, UNIFORM 3D 2MM X 2.5 CM/FCX140225, UNIFORM 3D 3MM X 3 CM/FCX140303, UNIFORM 3D 3MM X 4 CM/FCX140304, UNIFORM 3D 3MM X 6 CM/FCX140306, UNIFORM 3D 4MM X 6 CM/FCX140406, UNIFORM 3D 4MM X 8 CM/FCX140408, UNIFORM 3D 4MM X 4.5 CM/FCX140445, UNIFORM 3D 5MM X 7 CM/FCX140507, UNIFORM 3D 5MM X 12 CM/FCX140512, UNIFORM 3D 6MM X 9 CM/FCX140609, UNIFORM 3D 6MM X 15 CM/FCX140615, UNIFORM 3D 7MM X 11 CM/FCX140711, UNIFORM 3D 7MM X 17 CM/FCX140717, UNIFORM 3D 8MM X 12 CM/FCX140812, UNIFORM 3D 8MM X 20 CM/FCX140820, UNIFORM 3D 9MM X 14 CM/FCX140914, UNIFORM 3D 9MM X 22 CM/FCX140922, UNIFORM 3D 10MM X 16 CM/FCX141016, UNIFORM 3D 10MM X 25 CM/FCX141025, UNIFORM 3D 11MM X 18 CM/FCX141118, UNIFORM 3D 11MM X 27 CM/FCX141127, UNIFORM 3D 12MM X 19 CM/FCX141219, UNIFORM 3D 12MM X 30 CM/FCX141230, UNIFORM XL 5MM X 10 CM/FCX180510, UNIFORM XL 6MM X 12.2 CM/FCX180612, UNIFORM XL 7MM X 14.3 CM/FCX180714, UNIFORM XL 8MM X 13.6 CM/FCX180813, UNIFORM XL 8MM X 30 CM/FCX180830, UNIFORM XL 9MM X 15.3 CM/FCX180915, UNIFORM XL 9MM X 33 CM/FCX180933, UNIFORM XL 10MM X 17.1 CM/FCX181017, UNIFORM XL 10MM X 34 CM/FCX181034, UNIFORM XL 11MM X 18.9 CM/FCX181118, UNIFORM XL 11MM X 37 CM/FCX181137, UNIFORM XL 12MM X 20.7 CM/FCX181220, UNIFORM XL 12MM X 40 CM/FCX181240, UNIFORM XL 13MM X 43 CM/FCX181343, UNIFORM XL 14MM X 47 CM/FCX181447, UNIFORM XL 15MM X 50 CM/FCX181550, UNIFORM XL 16MM X 47 CM/FCX181647, UNIFORM XL 17MM X 50 CM/FCX181750, UNIFORM XL 18MM X 46 CM/FCX181846, UNIFORM XL 19MM X 50 CM/FCX181950, UNIFORM XL 20MM X 50 CM/FCX182050, FREEFORM MINI 1MM X 1CM/MCR091010, FREEFORM MINI 1MM X 1.5CM/MCR091015, FREEFORM MINI 1MM X 2CM/MCR091020, FREEFORM MINI 1MM X 2.5CM/MCR091025, FREEFORM MINI 1MM X 3CM/MCR091030, FREEFORM MINI 1MM X 4CM/MCR091040, FREEFORM MINI 1.5MM X 2CM/MCR091520, FREEFORM MINI 1.5MM X 2.5CM/MCR091525, FREEFORM MINI 1.5MM X 3CM/MCR091530, FREEFORM MINI 1.5MM X 4CM/MCR091540, FREEFORM MINI 2MM X 2CM/MCR092020, FREEFORM MINI 2MM X 3CM/MCR092030, FREEFORM MINI 2MM X 4CM/MCR092040, FREEFORM MINI 2MM X 6CM/MCR092060, FREEFORM MINI 2.5MM X 3.5CM/MCR092535, FREEFORM MINI 2.5MM X 4.5CM/MCR092545, FREEFORM MINI 2.5MM X 5.5CM/MCR092555, FREEFORM MINI 3MM X 4CM/MCR093040, FREEFORM MINI 3MM X 6CM/MCR093060, FREEFORM MINI 3MM X 8CM/MCR093080, FREEFORM 2MM X 2 CM/SCR120202, FREEFORM 3MM X 6 CM/SCR120306, FREEFORM 3MM X 8 CM/SCR120308, FREEFORM 4MM X 7 CM/SCR120407, FREEFORM 4MM X 10 CM/SCR120410, FREEFORM 4MM X 12 CM/SCR120412, FREEFORM 5MM X 10 CM/SCR120510, FREEFORM 5MM X 15 CM/SCR120515, FREEFORM 6MM X 10 CM/SCR120610, FREEFORM 6MM X 15 CM/SCR120615, FREEFORM 6MM X 20 CM/SCR120620, FREEFORM 7MM X 15 CM/SCR120715, FREEFORM 7MM X 21 CM/SCR120721, FREEFORM 8MM X 15 CM/SCR120815, FREEFORM 8MM X 24 CM/SCR120824, FREEFORM 9MM X 25 CM/SCR120925, FREEFORM 10MM X 30 CM/SCR121030, FREEFORM 12MM X 30 CM/SCR121230, FREEFORM 2.5MM X 5 CM/SCR122505, FREEFORM 2.5MM X 3.5 CM/SCR122535, FREEFORM 3.5MM X 9 CM/SCR123509, FREEFORM 3.5MM X 7.5 CM/SCR123575, HELIFORM SOFT 1.5MM X 4 CM/SHD100154, HELIFORM SOFT 2MM X 8 CM/SHD100208, HELIFORM SOFT 2.5MM X 10 CM/SHD100250, HELIFORM SOFT 2.5MM X 8 CM/SHD100258, HELIFORM SOFT 3MM X 8 CM/SHD100308, HELIFORM SOFT 3MM X 10 CM/SHD100310, HELIFORM SOFT 4MM X 8 CM/SHD100408, HELIFORM SOFT 4MM X 10 CM/ SHD100410, HELIFORM SOFT 5MM X 10 CM/ SHD100510, HELIFORM SOFT 5MM X 15 CM/ SHD100515, HELIFORM SOFT 6MM X 12 CM/ SHD100612, HELIFORM SOFT 6MM X 16 CM/ SHD100616, HELIFORM SOFT 7MM X 16 CM/ SHD100716, HELIFORM SOFT 7MM X 20 CM/ SHD100720, HELIFORM SOFT 8MM
Check your product
The notice names 23 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 10886704083986
- 10886704083979
- 10886704083962
- 10886704083993
- 10886704084013
- 10886704084006
- 10886704084037
- 10886704084020
- 10886704084044
- 10886704084051
- 10886704084068
- 10886704084075
- 10886704084082
- 10886704084099
- 10886704084167
- 10886704084174
- 10886704084181
- 10886704084198
- 10886704084211
- 10886704084228
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- UDI-D
Model or item number
On the rating label, the box or the instruction sheet.
- FCX100202
- UNIFORM 2.5MM X 3.3
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Cerenovus Inc — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Cerenovus Inc — 6303 Blue Lagoon Dr Ste 315, Miami, FL, 33126-6068
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Cerenovus Inc medical devices recalled?
Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention. Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 10886704083986, 10886704083979, 10886704083962; lot code UDI-D; model number FCX100202, UNIFORM 2.5MM X 3.3. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…
-
Class II · 11 days ago
Medtronic MiniMed, Inc. Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin recall: Some MiniMed Flex insulin pumps were released with missing (i.e
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed…
-
Class II · 11 days ago
Philips Medical Systems Spectral CT and Spectral CT 7500 RT recall: Software issues may affect the safety and performance of Spectral CT
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…
-
Class II · 11 days ago
Onkos Surgical Inc. My3D recall: Personalized pelvic implants inadvertently mixed up and distributed
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.