Class II U.S. Food and Drug Administration

C-Rad Positioning Ab Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/ recall: Due to issues with the system's stereotactic radiosurgery (SRS) treatment

Recall published 9 months ago

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

The hazard: Device safety concern

A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

What the notice says: Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-1092-2026
Published
January 21, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
C-Rad Positioning Ab
Product type
Medical devices
Units affected
189
Distribution
Nationwide
Made in
Sweden
Lot / code info
Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module

Products covered by this recall

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Check your product

The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

Model or item number

On the rating label, the box or the instruction sheet.

  • SP001-0026
  • SP003-0002 Software Version

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

C-Rad Positioning Ab — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. Report device problems to the FDA through MedWatch.
  3. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  4. Ask the supplier or clinic that provided it what the remedy is in your case.

Claiming on a recalled device

Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.

  • Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
  • Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
  • If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
  • Log the date, the name and the reference number of every call. Device recalls can run for months.

Company contact: C-RAD POSITIONING AB — Bredgrand 18, Uppsala

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was C-Rad Positioning Ab medical devices recalled?

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: model number SP001-0026, SP003-0002 Software Version. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.

Where was the recalled product sold?

It was distributed nationwide.

Related recalls