Percussionaire Corporation Phasitron 5, IPV In-Line Valve recall: IFU update; in-line valve, with pressure relief seal may leak when used
IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0031-2026
- Published
- October 22, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Percussionaire Corporation
- Product type
- Medical devices
- Units affected
- 59160
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- UDI-DI: 00849436000723. Lots: 240326, 250116, 240418, 250324, 240610, 250616, 240620, 241121, 230612, 240826, WO04884, 241118, WO06020, 241203, WO05019, 250111, WO04827, WO04756. Instructions for Use P20020 Rev F
Products covered by this recall
Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use
Check your product
The notice names 10 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 240326
- 250116
- 240418
- 250324
- 240610
- 250616
- 240620
- 241121
- 230612
Model or item number
On the rating label, the box or the instruction sheet.
- P5-TEE-20
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Percussionaire Corporation — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Percussionaire Corporation — 130 Mcghee Rd Ste 109, Sandpoint, ID, 83864-8409
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Percussionaire Corporation medical devices recalled?
IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure. IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: lot code 240326, 250116, 240418; model number P5-TEE-20. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 18 days ago
Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV HC recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV-2C recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation In-Line Valve, Product Number: P5-TEE recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Impulsator, Product Number: S00010, Phasitron 5 recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Travel Air, Product Number: S00055 recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV-1C recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…