Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV-2C recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-3322-2026
- Published
- October 7, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Percussionaire Corporation
- Product type
- Medical devices
- Units affected
- 229,430
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Product Number/UDI-DI: S00002-C/00849436000020, 00849436000563, 00849436000570, 00849436000143, 00849436000648; P5/00849436000501; Instruction Manual: P20046 Rev A
- UPC codes
- 00849436000020, 00849436000563, 00849436000570, 00849436000143, 00849436000648, 00849436000501
Products covered by this recall
Intrapulmonary Percussive Ventilation, IPV-2C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual
Check your product
The notice names 8 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00849436000020
- 00849436000563
- 00849436000570
- 00849436000143
- 00849436000648
- 00849436000501
Model or item number
On the rating label, the box or the instruction sheet.
- S00002-C
- UD
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Percussionaire Corporation — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Percussionaire Corporation — 130 Mcghee Rd Ste 109, Sandpoint, ID, 83864-8409
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Percussionaire Corporation medical devices recalled?
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00849436000020, 00849436000563, 00849436000570; model number S00002-C, UD. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 18 days ago
Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV HC recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation In-Line Valve, Product Number: P5-TEE recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Impulsator, Product Number: S00010, Phasitron 5 recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Travel Air, Product Number: S00055 recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV-1C recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…
-
Class II · 18 days ago
Percussionaire Corporation Intrapulmonary Percussive Ventilation, IPV 1 recall: Labeling updated warning not to use Intrapulmonary percussive ventilation
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV…