I.T.S. GmbH Brand Name: I.T.S. Pelvic Reconstruction System with the below recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1608-2026
- Published
- April 1, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- I.T.S. GmbH
- Product type
- Medical devices
- Units affected
- 46,637 units
- Distribution
- Nationwide
- Made in
- Austria
- Lot / code info
- 1. Article Number (Lot Numbers): 21218-16 (679/112166, 679/122229, 679/122230, 679/122304), 21219-16 (679/012441, 679/062456, 679/0624561, 679/112167, 679/122228). 2. Article Number (Lot Numbers): 21218-14(679/032417, 679/102231, 679/112158, 679/112164, 679/122231), 21219-14 (679/012505, 679/112159, 679/112165, 679/122227, 679/122305). 3. Article Number (Lot Numbers): 21216-10 (679/032501, 679/062453, 679/072276, 679/072510, 679/092203, 679/092401, 679/112168, 679/112409, 679/122270), 21216-12 (679/032503, 679/062344, 679/062454, 679/072277, 679/092204, 679/112170, 679/112413, 679/122271), 21217-10 (679/032502, 679/062346, 679/072278, 679/092402, 679/092404, 679/102216, 679/112169, 679/112423), 21217-12 (679/022238, 679/062345, 679/072513, 679/092403, 679/112171, 679/122407, 679/122426). 4. Article Number (Lot Numbers): 30591-36 (18/922090), 30591-70 (18/231539, 18/775273), 30591-75 (18/751307), 30591-80 (18/777295), 30591-85 (18/301308), 30591-90 (18/311309). 5. Article Number (Lot Numbers): 37422-12-N (18/145897, 18/461842), 37422-14-N (18/215281, 679/032001, 679/062320, 679/092355, 679/122326), 37422-16-N (679/052229, 679/052307, 679/102214, 679/102306, 679/122329), 37422-18-N (18/240520, 18/452529, 18/473354, 679/052316, 679/102215, 679/102337, 679/122330), 37422-20-N (679/0323147, 679/072306, 679/102211, 679/102338, 679/122331), 37422-22-N (18/092876, 18/543636, 679/012406, 679/062361, 679/072312, 679/102212, 679/102348), 37422-24-N (679/012407, 679/052333, 679/072313, 679/102349), 37422-26-N (679/012435, 679/062316, 679/072322, 679/102353), 37422-28-N (18/192499, 679/012422, 679/032211, 679/062317, 679/062323, 679/072323, 679/102361), 37422-30-N (18/217289, 679/012436, 679/032212, 679/0323142, 679/072324, 679/102354), 37422-32-N (18/218290, 679/012437, 679/052335, 679/052337, 679/072329, 679/112306), 37422-34-N (18/031830, 18/219291, 679/012446, 679/062337, 679/062341, 679/072330, 679/112318), 37422-36-N (18/220292, 18/240540, 18/734742, 679/012447, 679/042317, 679/042321, 679/062003, 679/072334, 679/102361), 37422-38-N (18/105881, 18/221293, 18/362296, 18/412482, 679/012243, 679/022407, 679/052331, 679/052334, 679/072349, 679/112323), 37422-40-N (18/052830, 18/222294, 18/240538, 18/493069, 18/736744, 679/022401, 679/032310, 679/042302, 679/072354, 679/112329), 37422-42-N (18/053831, 18/081544, 18/223295, 18/364298, 18/733745, 679/022402, 679/062371, 679/082304, 679/112330), 37422-44-N (18/054832, 18/224296, 18/365299, 18/737746, 18/781832, 679/022412, 679/042331, 679/042345, 679/082314, 679/112335), 37422-46-N (18/052578, 18/225314, 18/366300, 18/738747, 679/022419, 679/042313, 679/042318, 679/082319, 679/112336), 37422-48-N (18/056834, 18/226315, 679/022423, 679/052324, 679/052328, 679/072316, 679/092314, 679/112337), 37422-50-N (18/227316, 18/740748, 679/022436, 679/042336, 679/052314, 679/092342, 679/112338), 37422-55-N (18/113882, 18/368302, 679/032215, 679/052336, 679/062309, 679/092322, 679/112324), 37422-60-N (18/229318, 18/240517, 18/927095, 28/33034, 679/032403, 679/062349, 679/062351, 679/092325, 679/122307), 37422-65-N (18/118884, 18/230319, 18/271754, 18/743749, 18/761192, 28/35846, 679/032216, 679/072303, 679/092326, 679/122308), 37422-70-N (679/102213, 18/744750, 18/893916, 679/062368, 679/092338, 679/102213, 679/122309), 37422-75-N (18/122888, 18/231839, 679/072304, 679/122306), 37422-80-N (679/122254, 18/124887, 679/062352, 679/122320), 37422-85-N (18/056936, 18/271837, 679/012256, 679/092339), 37422-90-N (18/050600, 18/321273, 679/092340), 37422-95-N (18/130891, 679/092341), 37422-100-N (18/132892, 18/341607, 679/102403), 37422-105-N (679/102404), 37422-110-N (18/136894, 18/371843, 679/092337), 37422-115-N (18/138895, 18/391845, 679/012262), 37422-120-N (18/140896, 679/092349), 37592-18 (679/072351), 37592-20 (18/831792, 18/252862), 37592-22 (18/172773, 679/072356), 37592-24 (679/082301), 37592-26 (18/182774, 679/082303), 37592-28 (679/012140, 679/082310), 37592-30 (18/112775, 679/082317), 37592-32 (67…
Products covered by this recall
Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-Column Plate; Article Numbers: 21216-10, 21216-12, 21217-10, 21217-12. 4. Cancellous Screw; Article Numbers: 30591-36, 30591-70, 30591-75, 30591-80, 30591-85, 30591-90. 5. Cancellous Screw, Locking; Article Numbers: 37422-12-N, 37422-14-N, 37422-16-N, 37422-18-N, 37422-20-N, 37422-22-N, 37422-24-N, 37422-26-N, 37422-28-N, 37422-30-N, 37422-32-N, 37422-34-N, 37422-36-N, 37422-38-N, 37422-40-N, 37422-42-N, 37422-44-N, 37422-46-N, 37422-48-N, 37422-50-N, 37422-55-N, 37422-60-N, 37422-65-N, 37422-70-N, 37422-75-N, 37422-80-N, 37422-85-N, 37422-90-N, 37422-95-N, 37422-100-N, 37422-105-N, 37422-110-N, 37422-115-N, 37422-120-N, 37592-18, 37592-20, 37592-22, 37592-24, 37592-26, 37592-28, 37592-30, 37592-32, 37592-34, 37592-36, 37592-38, 37592-40, 37592-42, 37592-46, 37592-48, 37592-50, 37592-54, 37592-56, 37592-58, 37592-60, 37592-65, 37592-70, 37592-75, 37592-80, 37592-85, 37592-90. 6. Cannulated Screw, Cancellous; Article Numbers: 31851-60, 31851-65, 31851-70, 31851-75, 31851-80, 31851-85, 31851-90, 31851-95, 31851-100, 31851-105, 31851-110, 31851-115, 31851-120, 31851-125, 31851-130, 31851-135, 31851-140, 31851-145, 31851-150, 31851-155, 31851-160, 31851-165, 31851-170, 31851-175, 31851-180, 31851-185, 31851-190, 31851-195, 31851-200, 31854-60, 31854-65, 31854-70, 31854-75, 31854-80, 31854-85, 31854-90, 31854-95, 31854-100, 31854-105, 31854-110, 31854-115, 31854-120, 31854-125, 31854-130, 31854-135, 31854-140, 31854-145, 31854-150, 31854-155, 31854-160, 31854-165, 31854-170, 31854-175, 31854-180, 31854-185, 31854-190, 31854-195, 31854-200. 7. Cortical Screw; Article Numbers: 32351-10, 32351-12, 32351-14, 32351-16, 32351-18, 32351-20, 32351-22, 32351-24, 32351-26, 32351-28, 32351-30, 32351-32, 32351-34, 32351-36, 32351-38, 32351-40, 32351-42, 32351-44, 32351-46, 32351-48, 32351-50, 32351-55, 32351-60, 32351-65, 32351-70, 32351-75, 32351-80, 32351-85, 32352-16, 32352-18, 32352-20, 32352-22, 32352-24, 32352-26, 32352-28, 32352-30, 32352-32, 32352-34, 32352-36, 32352-38, 32352-40, 32352-42, 32352-44, 32352-46, 32352-48, 32352-50, 32352-55, 32352-60, 32352-65, 32352-70, 32352-75, 32352-80, 32352-85, 32352-90, 32352-95, 32352-100, 32352-105, 32352-110, 32352-115, 32352-120, 32455-18, 32455-20, 32455-22, 32455-24, 32455-26, 32455-28, 32455-30, 32455-32, 32455-34, 32455-36, 32455-38, 32455-40, 32455-42, 32455-44, 32455-46, 32455-48, 32455-50, 32455-52, 32455-54, 32455-56, 32455-58, 32455-60, 32455-65, 32455-70, 32455-75, 32455-80, 32455-85, 32455-90, 32475-32. 8. Cortical Screw, Locking; Article Numbers: 37455-18, 37455-20, 37455-22, 37455-24, 37455-26, 37455-28, 37455-30, 37455-32, 37455-34, 37455-36, 37455-38, 37455-40, 37455-42, 37455-44, 37455-46, 37455-48, 37455-50, 37455-52, 37455-54, 37455-56, 37455-60, 37455-65, 37455-70, 37455-75, 37455-80, 37455-85, 37455-90. 9. Curved Plate; Article Numbers: 211911-4, 211911-5, 211911-6, 211911-7, 211911-8, 211911-10, 211911-12, 211911-14, 211911-16, 211911-20. 10. Posterior Column Plate, Left; Article Numbers: 212121-6. 11. Posterior Column Plate, Right; Article Numbers: 212131-6. 12. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-SM. 13. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-LA. 14. Posterior Wall Plate, Extended, Right; Article Numbers: 212111-LA, 212111-SM. 15. Posterior Wall Plate, Left; Article Numbers: 212141-7, 212141-8. 16. Posterior Wall Plate, Right; Article Numbers: 212151-7, 212151-8. 17. PRS Plate; Article Numbers: 21194-6, 21194-7, 21194-8, 21195-10, 21195-12, 21195-7. 18. Quadrilateral Buttress Plate; Article Numbers: 21208-13, 21209-13. 19. Quadrilateral Buttress Plate; Article Numbers: 21208-11, 21209-11. 20. Rim Plate, Left; Article
What to do about Device
I.T.S. GmbH — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: I.T.S. GmbH — Autal 28, Lasnitzhohe
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was I.T.S. GmbH medical devices recalled?
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
How do I know if I have the recalled product?
The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Device enforcement record, and contact the company if you are unsure.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 7 months ago
I.T.S. GmbH HCS - Headless Compression Screw System, Headless Compression Screw recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH Olecranonplate with Angular Stability with the below descriptions recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH I.T.S. Screw System with the below descriptions and corresponding recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH HLS - Hand Locking Plates System with the below descriptions recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH Distal Humeral Plates with Angular Stability with the below recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…