Class II U.S. Food and Drug Administration

I.T.S. GmbH HLS - Hand Locking Plates System with the below descriptions recall: Updated MRI safety testing demonstrates higher RF-induced temperature

Recall published 7 months ago

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-1607-2026
Published
April 1, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
I.T.S. GmbH
Product type
Medical devices
Units affected
1,307 units
Distribution
Nationwide
Made in
Austria
Lot / code info
1. Article Number (Lot Numbers): 32151-6 (18/408462, 18/500495, 18/691266), 32151-8 (18/408866, 18/531370), 32151-9 (18/351668, 18/541373, 18/741269), 32151-10 (18/408986, 18/551374), 32151-11 (18/371670, 18/561375), 32151-12 (18/409087, 18/571376, 18/771272), 32151-13 (18/581377), 32151-14 (18/409289, 18/591378, 18/791274), 32151-15 (18/409390), 32151-16 (18/409491), 32151-18 (18/821277), 32151-20 (18/831278). 2. Article Number (Lot Numbers): 21505-4 (86/350493, 86/677353), 21506-4 (86/642718, 86/677157). 3. Article Number (Lot Numbers): 21507-4 (26/F65050), 21508-4 (26/F63934, 86/677180). 4. Article Number (Lot Numbers): 21517 (26/F66911). 5. Article Number (Lot Numbers): 21502 (86/677136), 21503-4 (86/337934), 21503-6 (86/868870), 21503-8 (26/F63936), 21503-10 (86/337896, 86/677146), 21504-6 (26/F63364, 86/868885), 21504-8 (26/F63365), 21504-10 (86/642711, 86/677154). 6. Article Number (Lot Numbers): 21513-3 (86/677385), 21513-5 (86/677389), 21514-3 (26/F63935), 21514-5 (86/642722, 86/677183, 86/869183). 7. Article Number (Lot Numbers): 37151-5 (18/561228), 37151-8 (18/166019), 37151-9 (18/625633), 37151-10 (18/626634), 37151-12 (18/628636, 18/671232), 37151-13 (18/166423), 37151-15 (18/206189), 37151-16 (18/132035, 18/691234), 37151-18 (18/142036), 37151-24 (18/167068, 18/172039), 37251-8 (18/421197), 37251-9 (18/591223), 37251-11 (18/619644), 37251-12 (18/620645, 18/962052), 37251-13 (18/521703), 37251-14 (18/531704, 18/951141), 37251-16 (18/002056, 18/031952, 18/623648, 18/971143), 37251-20 (18/571708, 26/F61731).

Products covered by this recall

HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape; Article Numbers: 21505-4, 21506-4. 3. HLS Plate, 3 T-Shape; Article Numbers: 21507-4, 21508-4. 4. HLS Plate, Square; Article Numbers: 21517. 5. HLS Plate, Straight; Article Numbers: 21502, 21503-4, 21503-6, 21503-8, 21503-10, 21504-6, 21504-8, 21504-10. 6. HLS Plate, Y-Shape; Article Numbers: 21513-3, 21513-5, 21514-3, 21514-5. 7. Locking Stabilization Screw; Article Numbers: 37151-5, 37151-8, 37151-9, 37151-10, 37151-12, 37151-13, 37151-15, 37151-16, 37151-18, 37151-24, 37251-8, 37251-9, 37251-11, 37251-12, 37251-13, 37251-14, 37251-16, 37251-20.

What to do about Device

I.T.S. GmbH — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If you are a facility, quarantine affected stock and log the correction.
  3. Report device problems to the FDA through MedWatch.
  4. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: I.T.S. GmbH — Autal 28, Lasnitzhohe

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was I.T.S. GmbH medical devices recalled?

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

How do I know if I have the recalled product?

The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Device enforcement record, and contact the company if you are unsure.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.