I.T.S. GmbH HLS - Hand Locking Plates System with the below descriptions recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1607-2026
- Published
- April 1, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- I.T.S. GmbH
- Product type
- Medical devices
- Units affected
- 1,307 units
- Distribution
- Nationwide
- Made in
- Austria
- Lot / code info
- 1. Article Number (Lot Numbers): 32151-6 (18/408462, 18/500495, 18/691266), 32151-8 (18/408866, 18/531370), 32151-9 (18/351668, 18/541373, 18/741269), 32151-10 (18/408986, 18/551374), 32151-11 (18/371670, 18/561375), 32151-12 (18/409087, 18/571376, 18/771272), 32151-13 (18/581377), 32151-14 (18/409289, 18/591378, 18/791274), 32151-15 (18/409390), 32151-16 (18/409491), 32151-18 (18/821277), 32151-20 (18/831278). 2. Article Number (Lot Numbers): 21505-4 (86/350493, 86/677353), 21506-4 (86/642718, 86/677157). 3. Article Number (Lot Numbers): 21507-4 (26/F65050), 21508-4 (26/F63934, 86/677180). 4. Article Number (Lot Numbers): 21517 (26/F66911). 5. Article Number (Lot Numbers): 21502 (86/677136), 21503-4 (86/337934), 21503-6 (86/868870), 21503-8 (26/F63936), 21503-10 (86/337896, 86/677146), 21504-6 (26/F63364, 86/868885), 21504-8 (26/F63365), 21504-10 (86/642711, 86/677154). 6. Article Number (Lot Numbers): 21513-3 (86/677385), 21513-5 (86/677389), 21514-3 (26/F63935), 21514-5 (86/642722, 86/677183, 86/869183). 7. Article Number (Lot Numbers): 37151-5 (18/561228), 37151-8 (18/166019), 37151-9 (18/625633), 37151-10 (18/626634), 37151-12 (18/628636, 18/671232), 37151-13 (18/166423), 37151-15 (18/206189), 37151-16 (18/132035, 18/691234), 37151-18 (18/142036), 37151-24 (18/167068, 18/172039), 37251-8 (18/421197), 37251-9 (18/591223), 37251-11 (18/619644), 37251-12 (18/620645, 18/962052), 37251-13 (18/521703), 37251-14 (18/531704, 18/951141), 37251-16 (18/002056, 18/031952, 18/623648, 18/971143), 37251-20 (18/571708, 26/F61731).
Products covered by this recall
HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape; Article Numbers: 21505-4, 21506-4. 3. HLS Plate, 3 T-Shape; Article Numbers: 21507-4, 21508-4. 4. HLS Plate, Square; Article Numbers: 21517. 5. HLS Plate, Straight; Article Numbers: 21502, 21503-4, 21503-6, 21503-8, 21503-10, 21504-6, 21504-8, 21504-10. 6. HLS Plate, Y-Shape; Article Numbers: 21513-3, 21513-5, 21514-3, 21514-5. 7. Locking Stabilization Screw; Article Numbers: 37151-5, 37151-8, 37151-9, 37151-10, 37151-12, 37151-13, 37151-15, 37151-16, 37151-18, 37151-24, 37251-8, 37251-9, 37251-11, 37251-12, 37251-13, 37251-14, 37251-16, 37251-20.
What to do about Device
I.T.S. GmbH — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: I.T.S. GmbH — Autal 28, Lasnitzhohe
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was I.T.S. GmbH medical devices recalled?
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
How do I know if I have the recalled product?
The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Device enforcement record, and contact the company if you are unsure.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 7 months ago
I.T.S. GmbH HCS - Headless Compression Screw System, Headless Compression Screw recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH Olecranonplate with Angular Stability with the below descriptions recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH I.T.S. Screw System with the below descriptions and corresponding recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH Distal Humeral Plates with Angular Stability with the below recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…
-
Class II · 7 months ago
I.T.S. GmbH I.T.S. Straight Plate with Angular Stability with the below recall: Updated MRI safety testing demonstrates higher RF-induced temperature
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the…