Fujirebio Diagnostics, Inc. Product Name: Lumipulse G pTau 217 Plasma Calibrators recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1303-2026
- Published
- February 11, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Fujirebio Diagnostics, Inc.
- Product type
- Medical devices
- Units affected
- 117 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model/Catalog Number: 81557; UDI-DI: 05414470815579; Lot Numbers: D5C6094U, D5C6125U;
- UPC codes
- 05414470815579
Products covered by this recall
Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The pTau 217 plasma concentration of a specimen is automatically calculated from the calibration curve, which is also automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G pTau 217 Plasma Calibrators: Liquid (Frozen), 1x1.5 mL (5 Concentrations) CAL 1 0 pg/mL pTau 217 Plasma Calibrator CAL 2 0.250 pg/mL pTau 217 Plasma Calibrator CAL 3 1.000 pg/mL pTau 217 Plasma Calibrator CAL 4 5.000 pg/mL pTau 217 Plasma Calibrator CAL 5 10.000 pg/mL pTau 217 Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Check your product
The notice names 6 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 05414470815579
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- D5C6094U
- D5C6125U
Model or item number
On the rating label, the box or the instruction sheet.
- 81557 Software Version
- 81557
Package size
Match the net weight or the piece count on the front of the pack.
- 1.5 mL
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Fujirebio Diagnostics, Inc. — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Fujirebio Diagnostics, Inc. — 201 Great Valley Pkwy, Malvern, PA, 19355-1308
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Fujirebio Diagnostics, Inc. medical devices recalled?
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result). Customers may experience inaccurate test results by observing a higher frequency of…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 05414470815579; lot code D5C6094U, D5C6125U; model number 81557 Software Version, 81557. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse pTau 217 Plasma Controls Model/ recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/ recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G pTau 217 Plasma Immunoreaction recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…