Fujirebio Diagnostics, Inc. Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1305-2026
- Published
- February 11, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Fujirebio Diagnostics, Inc.
- Product type
- Medical devices
- Units affected
- 1423 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model/Catalog Number: 235447; UDI-DI: 04987270235447; Lot Numbers: C5C6022U, C5C6045U;
- UPC codes
- 04987270235447
Products covered by this recall
Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology13 by a specific two-step immunoassay method on the LUMIPULSE G System. Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-¿-amyloid 1-42 monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-¿-amyloid 1-42 monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains detergents in 50 mM MES buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Check your product
The notice names 5 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 04987270235447
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- C5C6022U
- C5C6045U
Model or item number
On the rating label, the box or the instruction sheet.
- 235447 Software Version
- 235447
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Fujirebio Diagnostics, Inc. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Fujirebio Diagnostics, Inc. — 201 Great Valley Pkwy, Malvern, PA, 19355-1308
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Fujirebio Diagnostics, Inc. medical devices recalled?
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result). Customers may experience inaccurate test results by observing a higher frequency of…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 04987270235447; lot code C5C6022U, C5C6045U; model number 235447 Software Version, 235447. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse pTau 217 Plasma Controls Model/ recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G pTau 217 Plasma Calibrators recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/ recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…
-
Class II · 9 months ago
Fujirebio Diagnostics, Inc. Product Name: Lumipulse G pTau 217 Plasma Immunoreaction recall: Customers may experience inaccurate test results by observing a higher
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low…