Boston Scientific TheraSphere 360 Y-90 Management Platform recall: Software anomaly occurs when a user adds one or more vials to the cart
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
The hazard: Device malfunction
This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
What the notice says: Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.
Symptoms and when to call a doctor
If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-2771-2026
- Published
- July 29, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Boston Scientific Corporation
- Product type
- Medical devices
- Units affected
- 68 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)
- UPC codes
- 00191506068264
Products covered by this recall
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00191506068264
Model or item number
On the rating label, the box or the instruction sheet.
- M00139509001
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Malfunction
Boston Scientific Corporation — This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
- Check whether the remedy is a software update, and if so confirm afterwards in the device menu that the version actually changed.
- If you are a clinic or facility, quarantine affected stock and record the correction in your recall log.
- Report device problems to the FDA through MedWatch — that database is where this recall came from.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Boston Scientific Corporation — 1 Scimed Pl, Maple Grove, MN, 55311-1565
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Boston Scientific Corporation medical devices recalled?
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation. Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00191506068264; model number M00139509001. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class I · 25 days ago
Boston Scientific Corporation Imager II Angiographic Catheters M001314001 Imager recall: Tip may fracture or detach, leading to device fragments remaining within
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include…
-
Class II · last month
Boston Scientific Corporation StoneSmart Connect Console Contents recall: A subset of power adapters used with the connect console may exhibit
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result…
-
Class II · last month
Boston Scientific Corporation Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac recall: There is potential for a delay in AF Burden Alert notification affecting
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs)…
-
Class II · last month
Boston Scientific Corporation Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac recall: There is potential for a delay in AF Burden Alert notification affecting
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs)…
-
Class II · last month
Boston Scientific Corporation Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac recall: There is potential for a delay in AF Burden Alert notification affecting
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs)…
-
Class II · 2 months ago
Boston Scientific Corporation OverStitch Suture Cinch recall: Device may fail to deploy the cinch implant and/or fail to cut the suture
Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.