Boston Scientific Corporation OverStitch Suture Cinch recall: Device may fail to deploy the cinch implant and/or fail to cut the suture
Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2961-2026
- Published
- August 26, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Boston Scientific Corporation
- Product type
- Medical devices
- Units affected
- 60,654 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Material No. M00505530; GTIN: 00191506060046; Lot No. 36159783, 36179248, 36179249, 36179250, 36179674, 36181153, 36193585, 36195262, 36195281,36213130, 36213131, 36222433, 36223110, 36233350, 36233423, 36233424, 36238771, 36246477, 36247744, 36247745, 36257144, 36257145, 36264795, 36266604, 36285303, 36285304, 36291094, 36291142, 36296642, 36297718, 36298785, 36299374, 36307286, 36335098, 36336245, 36336246, 36336247, 36336248, 36336249, 36349543, 36358240, 36366410, 36368880, 36370038, 36376493, 36390716, 36400755, 36413832, 36429438, 36440653, 36440659, 36440798, 36441070, 36441390, 36443933, 36450076, 36451398, 36452061, 36454608, 36454861, 36461432, 36466525, 36472307, 36472448, 36472813, 36473087, 36475081, 36482317, 36482413, 36482837, 36482838, 36486869, 36488822, 36496658, 36498004, 36500343, 36500745, 36508877, 36509436, 36511012, 36519283, 36519622, 36564120, 36567088, 36567089, 36651483, 36651990, 36660007, 36670672, 36689082, 36691535, 36691536, 36692473, 36715146, 36715947, 36722429, 36733894, 36736063, 36737026, 36742475, 36747253, 36754086, 36756476, 36767158, 36770136, 36795290, 36802913, 36816032, 36816864, 36819466, 36829515, 36837613, 36840422, 36855429, 36859432, 36870323, 36884519, 36886399, 36887192, 36894183, 36906859, 37049911, 37053262, 37063221, 37071091, 37082225, 37256930, 37258435, 37265997, 37268058, 37276648, 37277853, 37332709, 37335148, 37341879, 37431423, 37462817, 37475144, 37479080, 37512600, 37516705, 37538409, 37614916, 37628854, 37701281, 37712590, 37747392, 37759135, 37768422, 37780435, 37804241, 37813732, 37824328, 37846636, 37861753, 37874884, 37906665, 37952534, 37976914, 37980916, 37987576, 38001581, 38020339, 38045267, 38055953, 38064514, 38087625, 38092029, 38119620, 38134479, 38149599, 38168976, 38265895, 38274202, 38277823, 38298897, 38307807, 38355532, 38404637, 38421970, 38438020, 38441891, 38448572, 38458999, 38480821, 38495024, 38498977, 38505659, 38526386, 38527090, 38540408, 38613546, 38620724, 38629370, 38672064, 38682255, 38692722, 38709195, 38711985, 38728857, 38738005, 38748320, 38759951, 38771565, 38787506, 38798542, 38808911, 38847193, 38866237, 38886764, 38917596, 38926776, 38936218, 38939941, 38948809, 39018825, 39029561, 39032429, 39043147, 39095582, 39114250, 39131778; Exp. Dat Range 25-Mar-2028 to 19-Apr-2029.
- UPC codes
- 00191506060046
Products covered by this recall
OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device.
Check your product
The notice names 9 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00191506060046
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 36159783
- 36179248
- 36179249
- 36179250
- 36179674
- 36181153
- 36193585
Package size
Match the net weight or the piece count on the front of the pack.
- 1 pk
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Boston Scientific Corporation — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Boston Scientific Corporation — 100 Boston Scientific Way, Marlborough, MA, 01752-1234
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Boston Scientific Corporation medical devices recalled?
Device may fail to deploy the cinch implant and/or fail to cut the suture as intended. Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00191506060046; lot code 36159783, 36179248, 36179249. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class I · 25 days ago
Boston Scientific Corporation Imager II Angiographic Catheters M001314001 Imager recall: Tip may fracture or detach, leading to device fragments remaining within
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include…
-
Class II · last month
Boston Scientific Corporation StoneSmart Connect Console Contents recall: A subset of power adapters used with the connect console may exhibit
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result…
-
Class II · last month
Boston Scientific Corporation Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac recall: There is potential for a delay in AF Burden Alert notification affecting
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs)…
-
Class II · last month
Boston Scientific Corporation Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac recall: There is potential for a delay in AF Burden Alert notification affecting
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs)…
-
Class II · last month
Boston Scientific Corporation Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac recall: There is potential for a delay in AF Burden Alert notification affecting
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs)…
-
Class II · 2 months ago
Boston Scientific Corporation Suture Cinch Long 1pk. Endoscopic tissue approximation recall: Device may fail to deploy the cinch implant and/or fail to cut the suture
Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.