Class II U.S. Food and Drug Administration

Boston Scientific Corporation Faradrive Steerable Sheath Clear IDE recall: Boston Scientific's complaint investigations have identified that a subset

Recall published 3 months ago

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-2754-2026
Published
July 29, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Boston Scientific Corporation
Product type
Medical devices
Units affected
400 units
Distribution
Nationwide
Made in
United States
Lot / code info
UDI-DI: 00810087181062; Lots: CL12910, CL12978, CL15153
UPC codes
00810087181062

Products covered by this recall

FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01

Check your product

The notice names 4 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00810087181062

Lot or batch code

Usually ink-stamped near the seam, the cap or the bottom of the pack.

  • CL12910
  • CL12978
  • CL15153 — 400 units

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Boston Scientific Corporation — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. Ask the supplier or clinic that provided it what the remedy is in your case.
  3. If you are a facility, quarantine affected stock and log the correction.
  4. Report device problems to the FDA through MedWatch.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: Boston Scientific Corporation — 4100 Hamline Ave N, Saint Paul, MN, 55112-5700

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Boston Scientific Corporation medical devices recalled?

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body. Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00810087181062; lot code CL12910, CL12978, CL15153 — 400 units. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.