Class II U.S. Food and Drug Administration

Bard Peripheral Vascular Inc Venclose digiRF Generators, Model Vcrfg1 recall: Software version 3.35 of the Venclose digiRF Generator incorporates

Recall published 1+ years ago

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

The hazard: Device malfunction

A malfunction was identified in this device. What matters most is whether your particular unit falls inside the affected model, lot or serial range.

What the notice says: Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.

Symptoms and when to call a doctor

If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-2638-2025
Published
October 8, 2025
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Bard Peripheral Vascular Inc
Product type
Medical devices
Units affected
1,725
Distribution
Nationwide
Made in
United States
Lot / code info
UDI-DI: 00858254006022. Generator Software Version 3.35

Products covered by this recall

Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).

Check your product

The notice names 1 code you can read straight off the package. Type the one printed on yours and the list below narrows to it.

Model or item number

On the rating label, the box or the instruction sheet.

  • VCRFG1

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Malfunction

Bard Peripheral Vascular Inc — A malfunction was identified in this device. What matters most is whether your particular unit falls inside the affected model, lot or serial range.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
  2. If you are a clinic or facility, quarantine affected stock and record the correction in your recall log.
  3. Report device problems to the FDA through MedWatch — that database is where this recall came from.
  4. If the device is implanted, or you depend on it daily, speak to your clinician before changing anything about how you use it.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: Bard Peripheral Vascular Inc — 1625 W 3rd St Bldg 1, Tempe, AZ, 85281-2438

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Bard Peripheral Vascular Inc medical devices recalled?

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: model number VCRFG1. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.