HeartSine Technologies Ltd recalls
3 recalls filed under HeartSine Technologies Ltd. The most recent was published October 7, 2026.
-
Class III · 18 days ago
HeartSine samaritan PAD recall: Products were distributed without the required Maintenance Supplement
Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.
-
Class II · last month
HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes recall: Due to a damaged (over stretched) membrane tail that causes an intermittent
Due to a damaged (over stretched) membrane tail that causes an intermittent fault.
-
Class II · 12 months ago
HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03 recall: Due to customer complaints, defibrillator pads may contain damage pins
Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
About this page
HeartSine Technologies Ltd appears here because a federal agency named it as the recalling firm or the brand on the label. A recall is not a judgement about everything the company sells — it covers only the lot codes and models listed in each notice.