Recalls published in October 2025
Everything the FDA, USDA FSIS, CPSC and NHTSA announced in October 2025, newest first.
-
Class II · 1+ years ago
B Braun Medical Inc Design Options, Perifix FX Epidural Anesthesia Tray. Product Code recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class II · 1+ years ago
B Braun Medical Inc Perifix FX Continuous Epidural Anesthesia Tray with Perifix 17 recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class II · 1+ years ago
Inpeco S.A. FlexLab X System. Potassium Test System. in vitro diagnostic recall: The FlexLab X System Sample Integrity Module (SIM) includes function
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This…
-
Class II · 1+ years ago
Corneat Vision, Ltd. EverPatch+, REF: 1044002, Permanent Scleral Reinforcement Matrix recall: Reviewed complaints have identified that conjunctival wound dehiscence
Reviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch.
-
Class II · 1+ years ago
B Braun Medical Inc Espocan Combined Spinal and Epidural Anesthesia Tray recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class II · 1+ years ago
ImaCor Inc. ClariTEE Miniaturized TEE probe recall: Production assembly error resulting in the potential for reversed
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex…
-
Class II · 1+ years ago
Applied Medical Technology Inc NutraGlide recall: The nasal feeding tubes, NutraGlide, subject to this recall have distal tips
The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces.
-
Class II · 1+ years ago
CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical recall: Reviewed complaints have identified that conjunctival wound dehiscence
Reviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch.
-
Class II · 1+ years ago
B Braun Medical Inc Perifix FX Continuous Epidural Anesthesia Tray with Perifix 17 recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class I · 1+ years ago
Abiomed, Inc. Automated Impella Controller , used for left heart support blood pump recall: Potential cybersecurity vulnerabilities related to the operating system
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
-
Class II · 1+ years ago
B Braun Medical Inc Combined Spinal and Epidural Anesthesia Tray recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class II · 1+ years ago
B Braun Medical Inc Combined Spinal and Epidural Anesthesia Tray 2 recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class II · 1+ years ago
GE Medical Systems, SCS AW Server 2.0, AW Server 3.0, AW Server 3.1, AW Server 3.2 recall: Firm has identified a security vulnerability in AW Server products
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity…
-
Class II · 1+ years ago
B Braun Medical Inc Perifix FX Continuous Epidural Anesthesia Tray with Perifix 17 recall: Potential for the lid of the catheter connector to be in the incorrect
Potential for the lid of the catheter connector to be in the incorrect position.
-
Class II · 1+ years ago
Canon Medical System, USA, INC. This device is a digital radiography/fluoroscopy recall: It has been found that when SPOT Fluoro is first turned
It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI…
-
Class I · 1+ years ago
CareFusion 303, Inc. BD Alaris Pump Infusion Sets for the following reference numbers recall: Infusion pump module used with compatible pump infusion sets may perform
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy…
-
Class I · 1+ years ago
NOxBOXi Nitric Oxide Delivery System recall: Fluctuations may be more likely to occur if the total flow through the device
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum…
-
Class II · 1+ years ago
GenoGenix LLC Ultra Burn for Injection, all strengths and presentations recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility
-
Class II · 1+ years ago
GenoGenix LLC Semaglutide with B12 for Injection, all strengths and presentations recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility
-
Class II · 1+ years ago
GenoGenix LLC Sterile Saline Reconstitution Solution for Injection recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility
-
Class II · 1+ years ago
B6 for Injections, all strengths and presentations, GenoGenix recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility
-
Class II · 1+ years ago
GenoGenix LLC Sermorelin Lyophilized powder for reconstitution, 10mg recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility
-
Class II · 1+ years ago
GenoGenix LLC BPC-157/Thymosin Beta-4 for Injection, a) 10/10mg, b) 15/15mg recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility
-
Class II · 1+ years ago
GenoGenix LLC Low Solubility Peptide Reconstitution Solution for Injection recall: Lack of Assurance of Sterility
Lack of Assurance of Sterility