NOxBOXi Nitric Oxide Delivery System recall: Fluctuations may be more likely to occur if the total flow through the device
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0028-2026
- Published
- October 15, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Noxbox Ltd
- Product type
- Medical devices
- Units affected
- 1667 units
- Distribution
- Tennessee
- Made in
- United Kingdom
- Lot / code info
- Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.
- UPC codes
- 05060541640009
Products covered by this recall
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 05060541640009
Serial number
On the rating plate, often next to the date of manufacture.
- s. — 1667 units
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Noxbox Ltd — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: NOXBOX LTD — Unit 1, Sittingbourne
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Noxbox Ltd medical devices recalled?
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz). Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 05060541640009. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Which states were affected?
The notice lists Tennessee.
Related recalls
-
Class II · 18 days ago
NOxBOXi Nitric Oxide Delivery System recall: Potential for irregular oxygen readings due to a manufacturing issue
Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
-
Class II · last month
NOxBOXi Nitric Oxide Delivery System recall: Potential for oxygen tubing leaks due to failure of the tubing or tubing
Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
-
Class I · 1+ years ago
NOxBOXi Nitric Oxide Delivery System recall: The unexpected internal fault / system diagnostic error to result
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without…
-
Class II · 1+ years ago
NOxBOXi Nitric Oxide Delivery System recall: Replacement of the internal Sample Pump due to high failure rates
Replacement of the internal Sample Pump due to high failure rates.
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…