Waldemar Link GmbH & Co. KG Endo-Model, Replacement Set, Tibial Plateau, right/left, L recall: The affected Endo-Model replacement set for a Rotating Hinge tibial plateau
The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2934-2026
- Published
- August 19, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Waldemar Link GmbH & Co. KG
- Product type
- Medical devices
- Units affected
- 1 US; 9 OUS
- Distribution
- Indiana
- Made in
- Germany
- Lot / code info
- UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069
- UPC codes
- 04026575432134
Products covered by this recall
Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16
Check your product
The notice names 5 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 04026575432134
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- for Implant REF 15-0027
- 16
Model or item number
On the rating label, the box or the instruction sheet.
- 15-0027
- 16
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Waldemar Link GmbH & Co. KG — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Waldemar Link GmbH & Co. KG (Mfg Site) — Ost Str. 4-10, Norderstedt
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Waldemar Link GmbH & Co. KG medical devices recalled?
The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant. The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 04026575432134; lot code for Implant REF 15-0027, 16; model number 15-0027, 16. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Which states were affected?
The notice lists Indiana.
Related recalls
-
Class II · 7 months ago
Waldemar Link GmbH & Co. KG Embrace Drill Tower, Wedged recall: The surgical technique for the Embrace Shoulder Instruments - Drill Tower has
The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However…
-
Class II · 7 months ago
Waldemar Link GmbH & Co. KG Embrace Drill Tower, Standard/Lateral recall: The surgical technique for the Embrace Shoulder Instruments - Drill Tower has
The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However…
-
Class II · 7 months ago
Waldemar Link GmbH & Co. KG Endo-Model Replacement Plateau recall: The potential for an implant with a plateau screw where the bushing
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
-
Class II · 7 months ago
Waldemar Link GmbH & Co. KG Endo-Model Replacement Plateau recall: The potential for an implant with a plateau screw where the bushing
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
-
Class II · 7 months ago
Waldemar Link GmbH & Co. KG Endo-Model SL Connection Component incl. PE-Plateau recall: The potential for an implant with a plateau screw where the bushing
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
-
Class II · 7 months ago
Waldemar Link GmbH & Co. KG Endo-Model Replacement Plateau recall: The potential for an implant with a plateau screw where the bushing
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.