Ventec Life Systems, Inc. V+Pro recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0129-2027
- Published
- October 28, 2026
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Ventec Life Systems, Inc.
- Product type
- Medical devices
- Units affected
- 39,568 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- All Serial Numbers: PRT-01185-000 -UDI: 00855573007877; PRT-01185-001 - UDI: 00850018761161; PRT-01185-002 - UDI: 00850018761154
- UPC codes
- 00855573007877, 00850018761161, 00850018761154
Products covered by this recall
V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.
Check your product
The notice names 7 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00855573007877
- 00850018761161
- 00850018761154
Model or item number
On the rating label, the box or the instruction sheet.
- PRT-01185-000
- PRT-01185-001
- PRT-01185-002
Serial number
On the rating plate, often next to the date of manufacture.
- PRT-01185-000
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Ventec Life Systems, Inc. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Ventec Life Systems, Inc. — 22002 26th Ave Se, Bothell, WA, 98021-4903
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Ventec Life Systems, Inc. medical devices recalled?
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00855573007877, 00850018761161, 00850018761154; model number PRT-01185-000, PRT-01185-001, PRT-01185-002. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class I · 5 days ago
Ventec Life Systems, Inc. VOCSN recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of…
-
Class I · 5 days ago
Ventec Life Systems, Inc. V+ Pro Emergency recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of…
-
Class I · 5 days ago
Ventec Life Systems, Inc. VC recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of…
-
Class I · 5 days ago
Ventec Life Systems, Inc. VCSN+Pro recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of…
-
Class I · 5 days ago
Ventec Life Systems, Inc. VOCSN recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of…
-
Class I · 5 days ago
Ventec Life Systems, Inc. VOCSN+Pro recall: Ventilator's Internal Flow transducer cable connector may disconnect
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of…