Class II U.S. Food and Drug Administration

Tandem Diabetes Care, Inc. t:slim X2 insulin pump with Interoperable Technology recall: A software defect in versions 7.9.0.1 and 7.10.1 of the pump software

Recall published 12 months ago

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

The hazard: Device malfunction

The notice describes a specific failure mode. That detail is what a clinician needs, because the right action differs sharply between a monitor, an implant and a test kit.

What the notice says: A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.

Symptoms and when to call a doctor

If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-0232-2026
Published
October 29, 2025
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Tandem Diabetes Care, Inc.
Product type
Medical devices
Units affected
122,958
Distribution
Nationwide
Made in
United States
Lot / code info
t:slim X2, UDI-DI: 00389152407012, 00389152407111, 00389152407210, 00389152407319 with software versions 7.9.0.1 and 7.10.1, User Guide versions not including and prior to: AW-1014562 Revision D; Tandem Mobi, UDI-DI: 00389152480114, 00389152481111, 00389152481319 with software version 7.9.0.1, User Guide versions not including and prior to: AW-1014938 Revision C
UPC codes
00389152407012, 00389152407111, 00389152407210, 00389152407319, 00389152480114, 00389152481111, 00389152481319

Products covered by this recall

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

Check your product

The notice names 14 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00389152407012
  • 00389152407111
  • 00389152407210
  • 00389152407319
  • 00389152480114
  • 00389152481111
  • 00389152481319

Model or item number

On the rating label, the box or the instruction sheet.

  • 1014070
  • 1014071
  • 1014072
  • 1014073
  • 1014801
  • 1014811
  • 1014813

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Malfunction

Tandem Diabetes Care, Inc. — The notice describes a specific failure mode. That detail is what a clinician needs, because the right action differs sharply between a monitor, an implant and a test kit.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
  2. If the device is implanted, or you depend on it daily, speak to your clinician before changing anything about how you use it.
  3. Check whether the remedy is a software update, and if so confirm afterwards in the device menu that the version actually changed.
  4. If you are a clinic or facility, quarantine affected stock and record the correction in your recall log.

Claiming on a recalled device

Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.

  • Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
  • Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
  • If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
  • Log the date, the name and the reference number of every call. Device recalls can run for months.

Company contact: Tandem Diabetes Care, Inc. — 12400 High Bluff Dr, San Diego, CA, 92130-3077

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Tandem Diabetes Care, Inc. medical devices recalled?

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia. A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00389152407012, 00389152407111, 00389152407210; model number 1014070, 1014071, 1014072. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.

Can I get my money back?

Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.

Where was the recalled product sold?

It was distributed nationwide.