Siemens Medical Solutions USA, Inc Magnetom Vida recall: There is a potential for an ice blockage to form or currently exist within
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
The hazard: High-powered magnet hazard
The problem with this product is magnets coming loose. A magnet ingestion looks like an ordinary stomach upset for the first day or two, which is exactly why these recalls are issued.
What the notice says: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Why it matters: Loose high-powered magnets are dangerous when more than one is swallowed. They attract each other through the walls of the intestine and pinch the tissue trapped between them, an injury that usually needs surgery.
Symptoms and when to call a doctor
If you think a child has swallowed magnets — even if they seem entirely well — go to an emergency department and say magnets are involved. Early symptoms are indistinguishable from a stomach bug: stomach pain, vomiting, poor appetite. A single magnet may pass on its own; two or more rarely do, and the team needs to know how many are unaccounted for.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-2667-2025
- Published
- October 8, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Siemens Medical Solutions USA, Inc
- Product type
- Medical devices
- Units affected
- 498 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model Number: 11060815. UDI Numbers: (01)04056869039176(21)176636, (01)04056869039176(21)177211, (01)04056869039176(21)175728, (01)04056869039176(21)175685, (01)04056869039176(21)175851, (01)04056869039176(21)175768, (01)04056869039176(21)177298, (01)04056869039176(21)177295, (01)04056869039176(21)176756, (01)04056869039176(21)175754, (01)04056869039176(21)177349, (01)04056869039176(21)176757, (01)04056869039176(21)177232, (01)04056869039176(21)175935, (01)04056869039176(21)176967, (01)04056869039176(21)175741, (01)04056869039176(21)177052, (01)04056869039176(21)176060, (01)04056869039176(21)176371, (01)04056869039176(21)177012, (01)04056869039176(21)177146, (01)04056869039176(21)175770, (01)04056869039176(21)176375, (01)04056869039176(21)176122, (01)04056869039176(21)176308, (01)04056869039176(21)175669, (01)04056869039176(21)176212, (01)04056869039176(21)176755, (01)04056869039176(21)176781, (01)04056869039176(21)176302, (01)04056869039176(21)176036, (01)04056869039176(21)175766, (01)04056869039176(21)176825, (01)04056869039176(21)177151, (01)04056869039176(21)176087, (01)04056869039176(21)176771, (01)04056869039176(21)177299, (01)04056869039176(21)176451, (01)04056869039176(21)177352, (01)04056869039176(21)177354, (01)04056869039176(21)176779, (01)04056869039176(21)175870, (01)04056869039176(21)175838, (01)04056869039176(21)175692, (01)04056869039176(21)177359, (01)04056869039176(21)177119, (01)04056869039176(21)175977, (01)04056869039176(21)176037, (01)04056869039176(21)177244, (01)04056869039176(21)176686, (01)04056869039176(21)175831, (01)04056869039176(21)177096, (01)04056869039176(21)176660, (01)04056869039176(21)177034, (01)04056869039176(21)176639, (01)04056869039176(21)176397, (01)04056869039176(21)175805, (01)04056869039176(21)176358, (01)04056869039176(21)176293, (01)04056869039176(21)176118, (01)04056869039176(21)175681, (01)04056869039176(21)177088, (01)04056869039176(21)177035, (01)04056869039176(21)176093, (01)04056869039176(21)176189, (01)04056869039176(21)177337, (01)04056869039176(21)177333, (01)04056869039176(21)176094, (01)04056869039176(21)176736, (01)04056869039176(21)177105, (01)04056869039176(21)177007, (01)04056869039176(21)175787, (01)04056869039176(21)177381, (01)04056869039176(21)176027, (01)04056869039176(21)176340, (01)04056869039176(21)177144, (01)04056869039176(21)176258, (01)04056869039176(21)176363, (01)04056869039176(21)176561, (01)04056869039176(21)176665, (01)04056869039176(21)176577, (01)04056869039176(21)176572, (01)04056869039176(21)176559, (01)04056869039176(21)176696, (01)04056869039176(21)176452, (01)04056869039176(21)176081, (01)04056869039176(21)175853, (01)04056869039176(21)175893, (01)04056869039176(21)177033, (01)04056869039176(21)177392, (01)04056869039176(21)177116, (01)04056869039176(21)177124, (01)04056869039176(21)175897, (01)04056869039176(21)175883, (01)04056869039176(21)177329, (01)04056869039176(21)176567, (01)04056869039176(21)175640, (01)04056869039176(21)176033, (01)04056869039176(21)176316, (01)04056869039176(21)176529, (01)04056869039176(21)176219, (01)04056869039176(21)176254, (01)04056869039176(21)176865, (01)04056869039176(21)176709, (01)04056869039176(21)176059, (01)04056869039176(21)175829, (01)04056869039176(21)175706, (01)04056869039176(21)176635, (01)04056869039176(21)175903, (01)04056869039176(21)176374, (01)04056869039176(21)176217, (01)04056869039176(21)176832, (01)04056869039176(21)175834, (01)04056869039176(21)175655, (01)04056869039176(21)175660, (01)04056869039176(21)176630, (01)04056869039176(21)177258, (01)04056869039176(21)177162, (01)04056869039176(21)176661, (01)04056869039176(21)177371, (01)04056869039176(21)176829, (01)04056869039176(21)175907, (01)04056869039176(21)176956, (01)04056869039176(21)177213, (01)04056869039176(21)176570, (01)04056869039176(21)176532, (01)04056869039176(21)177257, (01)04056869039176(21)177262, (01)04056869039176(21)176319, (01)04056869039176(21)175850, (01)04056869039176(21)176332, (01)04056869039176(21)176096, (0…
- UPC codes
- 04056869039176
Products covered by this recall
MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 04056869039176
Model or item number
On the rating label, the box or the instruction sheet.
- 11060815. 510
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Magnets
Siemens Medical Solutions USA, Inc — The problem with this product is magnets coming loose. A magnet ingestion looks like an ordinary stomach upset for the first day or two, which is exactly why these recalls are issued.
- Put the product out of reach and count the magnets against the number the instructions say it should have.
- If any are unaccounted for and a child could have reached them, go to an emergency department and tell them magnets may have been swallowed.
- Make sure older children know that using magnets as fake piercings is a recognised route to a serious injury.
- Follow the remedy described on this page rather than keeping the item with the magnets glued back in.
If the claim stalls
Most recall claims are paid without argument. The ones that stall usually stall for a reason you can fix.
- Check you used the recall channel rather than ordinary customer support — the two queues rarely talk to each other.
- Paid by card? A chargeback for goods that are unfit for sale is a legitimate fallback once the company has had a fair chance.
- Send the photos they asked for in the format they asked for. A missing date code is the single commonest reason a claim sits unprocessed.
- Give it ten working days, then chase in writing and quote your claim reference.
Company contact: Siemens Medical Solutions USA, Inc — 40 Liberty Blvd, Malvern, PA, 19355-1418
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Siemens Medical Solutions USA, Inc medical devices recalled?
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 04056869039176; model number 11060815. 510. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Put the product out of reach and count the magnets against the number the instructions say it should have.
Can I get my money back?
Check you used the recall channel rather than ordinary customer support — the two queues rarely talk to each other.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 4 months ago
Siemens Medical Solutions USA, Inc ARTIS icono ceiling recall: Potential issue with the Display ceiling suspension (DCS Fix XL) that holds
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor…
-
Class II · 4 months ago
Siemens Medical Solutions USA, Inc ARTIS icono floor recall: Potential for loss of movement capability after adjusting the detector lift
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
-
Class II · 4 months ago
Siemens Medical Solutions USA, Inc ARTIS icono biplane recall: Potential for loss of movement capability after adjusting the detector lift
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
-
Class II · 4 months ago
Siemens Medical Solutions USA, Inc ARTIS icono ceiling recall: Potential for loss of movement capability after adjusting the detector lift
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
-
Class II · 5 months ago
Siemens Medical Solutions USA, Inc Artis Icono biplane recall: During patient examinations, the system may sporadically display a 0 x-ray
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed…
-
Class II · 5 months ago
Siemens Medical Solutions USA, Inc ARTIS pheno, Model 10849000. interventional recall: During patient examinations, the system may sporadically display a 0 x-ray
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed…