Siemens Healthcare Diagnostics Inc epoc BGEM BUN Test Card for the epoc Blood recall: Use of the affected product can result in discrepant low pH and measured
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2005-2026
- Published
- May 6, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Siemens Healthcare Diagnostics Inc
- Product type
- Medical devices
- Units affected
- 71725 (38025 US, 33700 OUS)
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);
- UPC codes
- 00809708121860
Products covered by this recall
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Check your product
The notice names 6 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00809708121860
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 01-26024-70
- 01-26026-70
- 01-26028-70
- 01-26030-70
Package size
Match the net weight or the piece count on the front of the pack.
- 25 pk
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Siemens Healthcare Diagnostics Inc — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Siemens Healthcare Diagnostics Inc — 2 Edgewater Dr, Norwood, MA, 02062-4637
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Siemens Healthcare Diagnostics Inc medical devices recalled?
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter. Use of the affected product can result in discrepant…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00809708121860; lot code 01-26024-70, 01-26026-70, 01-26028-70. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 18 days ago
Siemens Healthcare Diagnostics, Inc. Atellica CH Revised C-Reactive Protein recall: The potential for falsely elevated results
The potential for falsely elevated results.
-
Class II · 2 months ago
Siemens Healthcare Diagnostics, Inc. Beckman Coulter recall: Due to unacceptable precision performance exhibited in certain lots
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated…
-
Class II · 2 months ago
Siemens Healthcare Diagnostics, Inc. Syva EMIT 2000 Quinidine Assay REF: 4Q019UL recall: Due to unacceptable precision performance exhibited in certain lots
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated…
-
Class III · 2 months ago
Siemens Healthcare Diagnostics, Inc. Syva Emit tox Serum Tricyclic Antidepressant recall: out-of-specification sodium azide concentration in Syva Emit Tox Serum
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results…
-
Class II · 2 months ago
Siemens Healthcare Diagnostics, Inc. Atellica CH Urine Albumin REF: 11537225 recall: Processing of urine albumin assay on the CI analyzers may lead
Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results…
-
Class III · 2 months ago
Siemens Healthcare Diagnostics, Inc. Syva Emit Disopyramide Assay Enzyme Immunoassay recall: Out-of-specification sodium azide concentration in Syva Emit Disopyramide
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not…