Raysearch Laboratories Ab RayStation with the following product descriptions: 1 recall: DICOM SOP Instance UID and Series Instance UID from RayGateway
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2494-2025
- Published
- September 10, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Raysearch Laboratories Ab
- Product type
- Medical devices
- Units affected
- 0
- Distribution
- Nationwide
- Made in
- Sweden
- Lot / code info
- 1. Software Version: 6.1.1.2, UDI: 0735000201008220170529. 2. Software Version: 6.2.0.7, UDI: 0735000201007520170630. 3. Software Version: 6.3.0.6, UDI: 0735000201024220190923. Expiration Date: 2024-10-18
Products covered by this recall
RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1, Software Version: 6.1.1.2. 2. RayStation 6 Service Pack 2, Software Version: 6.2.0.7. 3. RayStation 6 Service Pack 3, Software Version: 6.3.0.6. Product Description: Radiation Therapy Treatment Planning System.
Check your product
The notice names 1 code you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Best-by / use-by date
Printed beside the lot code, not on the price sticker.
- 2024-10-18 — 0
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Raysearch Laboratories Ab — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: RAYSEARCH LABORATORIES AB — Eugeniavagen 18c, Stockholm
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Raysearch Laboratories Ab medical devices recalled?
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: best-by date 2024-10-18 — 0. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 4 months ago
Raysearch Laboratories Ab RayCare 2024A SP1. Oncology Information Systems recall: Beam set delivery note issues in which there is potential data loss
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is…
-
Class II · 4 months ago
Raysearch Laboratories Ab RayCare 2024A SP4. Oncology Information Systems recall: Beam set delivery note issues in which there is potential data loss
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is…
-
Class II · 9 months ago
Raysearch Laboratories Ab Brand Name: RayStation Product Name: RayStation/RayPlan recall: Potential for the invalidation for calculated radiation dose does not work
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for…
-
Class II · 9 months ago
Raysearch Laboratories Ab Brand Name: RayStation Product Name: RayStation/RayPlan recall: Potential for the invalidation for calculated radiation dose does not work
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for…
-
Class II · 9 months ago
Raysearch Laboratories Ab Brand Name: RayStation Product Name: RayStation/RayPlan recall: Potential for the invalidation for calculated radiation dose does not work
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for…
-
Class II · 9 months ago
Raysearch Laboratories Ab Brand Name: RayStation Product Name: RayStation/RayPlan recall: Potential for the invalidation for calculated radiation dose does not work
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for…