Physio-Control, Inc. Lifepak 15, REF: 99577-000025, 99577-000047 recall: Some defibrillators serviced between July 2023 through November 2025
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2065-2026
- Published
- May 13, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Physio-Control, Inc.
- Product type
- Medical devices
- Units affected
- 206
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Device/UDI/Serial Number: LIFEPAK 15/0088387391164815011543166573, 0088387387169015033143394891, 0088387387169015050743524360, 0088387387169015092843914574, 0088387387204815093043915122, 0088387387169016010444187447, 0088387387162116012044252611, 0088387387169016012644262721, 0088387387169016012644277082, 0088387387169016012744282280, 0088387387171316060244574340, 0088387387171316080444896329, 0088387387169016083144962804, 0088387387170617013145447229, 0088387387169017041245743786, 0088387387171317050945834209, 0088387387171317062845998142, 0088387387171317082846103043, 0088387387169017101046303288, 0088387387171317103146303092, 0088387387162117122146587130, 0088387387171317122646602003, 0088387387171317122846608783, 0088387387171317122946614788, 0088387387171318022246813507, 0088387387169018032946966267, 0088387387171318061947337396, 0088387387169018070547401928, 0088387387171318092747611441, 0088387387169018112947784931, 0088387387169018112947786317, 0088387387171318121847847661, 0088387387171318121847848950, 0088387387708118121947845317, 0088387387708118121947855371, 0088387387169018122047857172, 0088387387169018122047858660, 0088387387171318122847872868, 0088387387171319031248018370, 0088387387169019061248247559, 0088387387168319062748298704, 0088387387169019100848453060, 0088387387171319101448455890, 0088387387171319121848586351, 0088387387171319121948586599, 0088387387168320012848651507, 0088387387171320030548701689, 0088387387708120042848784975, 0088387387171320071648887706, 0088387387168320081748921846, 0088387387169020092448981398, 0088387387169020102649032288, 0088387387169020102649050140, 0088387387162120110249055931, 0088387387168320122949156402, 0088387387169021011149162067, 0088387387169021011149167463, 0088387387169021013049212578, 0088387387169021020949225869, 0088387387169021020949226060, 0088387387176821021949240566, 0088387387168321033149345530, 0088387387169021042149378981, 0088387387169021042649381781, 0088387387169021042749381576, 0088387387168321081849559854, 0088387387168321081949560127, 0088387387164521091649607224, 0088387387171321102149662806, 0088387377697122070750045009, 0088387387169022090150136719, 0088387387171322090850152289, 0088387387708122091350158293, 0088387387169022101950223934, 0088387387708122112150288760, 0088387387171322120150307353, 0088387387169023062850629377, 0088387387169023072450653772, 0088387387169023072450667197, 0088387387708123072850669746, 0088387387708123082850714419, 0088387387171323082950720588, 0088387387169023112150850129, 0088387387708123122050883524, 0088387388434824052251092313, 0088387387170624121151422867, 0088387387170624122351440213, 0088387387170624122351444039, 0088387388434824052251092313, 0088387387169019100848453060, 0088387387201715090843839264, 0088387387171318121847850793/ 38523489, 39736196, 40465345, 40717931, 40765749, 41130420, 41139184, 41420681, 41446039, 41446693, 41704583, 42221131, 42920956, 42921954, 43166573, 43394891, 43524360, 43914574, 43915122, 44187447, 44252611, 44262721, 44277082, 44282280, 44574340, 44896329, 44962804, 45447229, 45743786, 45834209, 45998142, 46103043, 46303288, 46303092, 46587130, 46602003, 46608783, 46614788, 46813507, 46966267, 47337396, 47401928, 47611441, 47784931, 47786317, 47847661, 47848950, 47845317, 47855371, 47857172, 47858660, 47872868, 48018370, 48247559, 48298704, 48453060, 48455890, 48586351, 48586599, 48651507, 48701689, 48784975, 48887706, 48921846, 48981398, 49032288, 49050140, 49055931, 49156402, 49162067, 49167463, 49212578, 49225869, 49226060, 49240566, 49345530, 49378981, 49381781, 49381576, 49559854, 49560127, 49607224, 49662806, 50045009, 50136719, 50152289, 50158293, 50223934, 50288760, 50307353, 50629377, 50653772, 50667197, 50669746, 50714419, 50720588, 50850129, 50883524, 51092313, 51422867, 51440213, 51444039, 51092313, 48453060, 43839264, 47850793, 38369584. LIFEPAK 1000/0088507469887815020943235807, 0088507469895315031343346860, 0088507469887815111944068667, 0088507469887815111944068669,…
Products covered by this recall
LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122
Check your product
The notice names 13 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Model or item number
On the rating label, the box or the instruction sheet.
- 99577-000025
- 99577-000047
- 99577-001256
- 99577-001
- 99425-000025
- 99425-000023
- 99425-000096
- 99335-000013
- 99335-000031
- 99507-000122 — Device
- UDI
Serial number
On the rating plate, often next to the date of manufacture.
- LIFEPAK 15
- 0088387391164815011543166573
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Physio-Control, Inc. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Physio-Control, Inc. — 11811 Willows Rd Ne, Redmond, WA, 98052-2003
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Physio-Control, Inc. medical devices recalled?
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy. Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: model number 99577-000025, 99577-000047, 99577-001256. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 7 months ago
Physio-Control, Inc. Physio Control Infant/Child Reduced Energy Defibrillator recall: Due to pediatric defibrillator electrode delamination
Due to pediatric defibrillator electrode delamination
-
Class II · 1+ years ago
Physio-Control, Inc. Lifepak CR2 Part / Catalog Numbers / Udi-Di code: CR2-2-001451 recall: Due to required inspections not being performed on products/units that have
Due to required inspections not being performed on products/units that have gone through servicing.
-
Class II · 1+ years ago
Physio-Control, Inc. Lifepak 15 Part / Catalog Numbers / Udi-Di code: V15-7-000022 / recall: Due to required inspections not being performed on products/units that have
Due to required inspections not being performed on products/units that have gone through servicing.
-
Class II · 1+ years ago
Physio-Control, Inc. Lifepak 1000 Part / Catalog Numbers / Udi-Di code: 320371500023 / recall: Due to required inspections not being performed on products/units that have
Due to required inspections not being performed on products/units that have gone through servicing.
-
Class II · 11 days ago
Stryker Sustainability Solutions Bard EP XT Decapolar Steerable recall: Diagnostic electrophysiology (EP) catheter may have a breached sterile
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could…
-
Class II · 11 days ago
Steris Corporation Verify Biological Steam Test Pack Model/ recall: Geobacillus stearothermophilus spore population in affected products may fall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam…