Philips Respironics Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal recall: The use of in-line nebulizers placed in certain locations may lead to aerosol
The use of in-line nebulizers placed in certain locations may lead to aerosol deposits forming over time on the ventilator flow sensor. Should this occur, there is a possibility of inaccurate flow measurements affecting therapy.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Philips Respironics has determined that in some circumstances, the use of in-line nebulizers placed in certain locations may lead to aerosol deposits forming over time on the ventilator flow sensor.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Published
- October 7, 2024
- Risk level
- Elevated risk
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Respironics
- Product type
- Medical devices
Products covered by this recall
Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 ventilators
What the company announced
Additional Usage Instructions issued by Philips as voluntary correction (Urgent Medical Device Correction in the US, Field Safety Notice outside of the US)
Additional Usage Instructions provide user action and guidance for continued use of affected devices
Philips Respironics has not received any specific complaints of device malfunctions resulting from in-line nebulizer use; issue identified by retrospective review of flow sensor performance
Philips Respironics
Philips Respironics issued a voluntary correction for Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 ventilators regarding the use of in-line nebulizers.
An Urgent Medical Device Correction (Field Safety Notice in the rest of world) is being sent to health systems and equipment distributors and is being communicated through this additional update to ensure patients who use the devices at home are informed in a timely manner.
Patients will receive communication from their usual health provider and should not discontinue therapy without consulting their healthcare provider. This update, and the associated Urgent Medical Device Correction (Field Safety Notice in the rest of world) provides instructions for user action and guidance for continued use of affected devices. Philips Respironics is investigating a solution and will provide additional information when available.
Philips Respironics has not received any specific complaints of device malfunctions resulting from in-line nebulizer use. As part of ongoing quality management, the company performed a retrospective review and identified a potential issue with flow sensors. At the time of this notification, no patient deaths have been reported.
An in-line nebulizer is a commonly-used accessory that converts therapeutic liquids (such as medication/saline solutions) into an aerosol to treat different respiratory symptoms.
Philips Respironics has determined that in some circumstances, the use of in-line nebulizers placed in certain locations may lead to aerosol deposits forming over time on the ventilator flow sensor. Should this occur, there is a possibility of inaccurate flow measurements affecting therapy.
Trilogy Evo ventilators that have never been used with in-line nebulizers are not affected by this issue and can continue to be used in accordance with the guidance in the field safety notice.
Potential patient impacts
As indicated in the Urgent Medical Device Correction (Field Safety Notice in the rest of world), if aerosol deposits accumulate over time on the flow sensor, there is a possibility of patient impact due to one or more of the following conditions: the ventilator may become inoperative following stand by or powering off, leading to delay in therapy; and over-delivery of inhaled and exhaled air (tidal volume) and/or under-delivery of oxygen for devices with a specific concentration setting (FiO2).
Potential hazards to patients from these conditions may include: respiratory discomfort; lung injury from too much air in the lungs or change in pressure (volutrauma/barotrauma); low oxygen saturation; or shortness of breath (dyspnea).
Customer / user immediate actions required
All Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 users, regardless of in-line nebulizer use are advised to take the following steps from the device Instructions for Use:
Ensure that specific alarms are set appropriately for patients.
Ensure that alternate sources of ventilation are available in case a ventilator ceases to function.
If using a Trilogy Evo O2, Trilogy Evo Universal, or Trilogy EV300 device with a specific oxygen concentration (FiO2) setting:
Ensure patients receive adequate oxygenation by continuous pulse oximetry (SpO2) monitoring in accordance with treatment protocols.
Use an external FiO2 analyzer in specific patient cases, and switch to an alternative ventilator if an external FiO2 analyzer is not available.
As indicated in the Instructions for use, maintain an immediately available back‐up device or alternative ventilator for rapid therapy transition.
If using in-line nebulizer treatments:
Follow the visual instructions in the field safety notice for correct nebulizer placement.
For specific prescriptions indicated in the field safety notice, transition patient to alternate device configurations.
Customers who require further information or support concerning this issue, may contact their local Philips Respironics Customer Service at:
For DMEs and Homecare customers: 1-800-345-6443, press option 4, then option 5
For Hospital customers: 1 (800) 722-9377, press option 2
Any adverse events experienced with the use of this product should be reported to the FDA’s MedWatch Program by phone at 1-800-FDA-1088, by fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch Web site at www.fda.gov/medwatch.
Reproduced from the company announcement posted on fda.gov. Links and contact forms in the original are not repeated here; the official notice below has them.
What to do about Device
Philips Respironics — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Philips Respironics, Inc. — Consumers: Philips Respironics Customer Service, 800-345-6443 — Media: Mario Fante, 603-560-9226, [email protected]
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Company recall announcements record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Respironics medical devices recalled?
The use of in-line nebulizers placed in certain locations may lead to aerosol deposits forming over time on the ventilator flow sensor. Should this occur, there is a possibility of inaccurate flow measurements affecting therapy. Philips Respironics has determined that in some circumstances, the use of in-line nebulizers placed in certain locations may lead to aerosol deposits forming over time on the ventilator flow sensor.
How do I know if I have the recalled product?
The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Company recall announcements record, and contact the company if you are unsure.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Related recalls
-
Class I · 7 months ago
Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00 recall: In some situations, the Obstruction Alarm does not trigger within
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
-
Class I · 7 months ago
Philips Respironics Trilogy Evo Universal recall: In some situations, the Obstruction Alarm does not trigger within
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
-
Class I · 7 months ago
Philips Respironics Trilogy EV300, Software Version 1.05.15.00 recall: In some situations, the Obstruction Alarm does not trigger within
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
-
Class I · 7 months ago
Philips Respironics Trilogy Evo, Software Version 1.05.15.00 recall: In some situations, the Obstruction Alarm does not trigger within
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
-
Class I · 7 months ago
Philips Respironics Trilogy Evo Universal recall: Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume…
-
Class I · 7 months ago
Philips Respironics Trilogy EV300, Software Version 1.05.15.00 recall: Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume…