Philips North America Cardiac Workstation 7000 recall: Failure of Environmental Stress Testing from a pinched power module wire
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0166-2026
- Published
- October 22, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips North America
- Product type
- Medical devices
- Units affected
- 29 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model Number: 860441; UDI-DI: 00884838094833; Serial Numbers: US42520284, US32520248, US32520244, US22520238, US22520221;
- UPC codes
- 00884838094833
Products covered by this recall
Cardiac Workstation 7000; Model Number: 860441;
Check your product
The notice names 6 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838094833
Model or item number
On the rating label, the box or the instruction sheet.
- 860441
Serial number
On the rating plate, often next to the date of manufacture.
- US42520284
- US32520248
- US32520244
- US225
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips North America — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Philips North America — 222 Jacobs St, Cambridge, MA, 02141-2296
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips North America medical devices recalled?
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short. Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838094833; model number 860441. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 18 days ago
Philips North America Ingenia 3.0T and 3.0T CX recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips North America MR 7700 recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips North America Ingenia 1.5T, 1.5T CX, and 1.5T S recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips North America Evolution upgrade 1.5T amd Evolution upgrade 3.0T recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips Medical Systems Technologies Ltd. Spectral CT 7500 recall: The Bolus tracking scan application may crash with a system error message
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold…
-
Class II · 18 days ago
Philips North America Upgrade dStream recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…