Philips North America Cardiac Workstation 5000 recall: Failure of Environmental Stress Testing from a pinched power module wire
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0165-2026
- Published
- October 22, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips North America
- Product type
- Medical devices
- Units affected
- 28 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Model Number: 860439; UDI-DI: 00884838094826; Serial Numbers:
- UPC codes
- 00884838094826
Products covered by this recall
Cardiac Workstation 5000; Model Number: 860439;
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838094826
Model or item number
On the rating label, the box or the instruction sheet.
- 860439
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips North America — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Philips North America — 222 Jacobs St, Cambridge, MA, 02141-2296
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips North America medical devices recalled?
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short. Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838094826; model number 860439. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 18 days ago
Philips North America Ingenia 3.0T and 3.0T CX recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips North America MR 7700 recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips North America Ingenia 1.5T, 1.5T CX, and 1.5T S recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips North America Evolution upgrade 1.5T amd Evolution upgrade 3.0T recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…
-
Class II · 18 days ago
Philips Medical Systems Technologies Ltd. Spectral CT 7500 recall: The Bolus tracking scan application may crash with a system error message
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold…
-
Class II · 18 days ago
Philips North America Upgrade dStream recall: When exporting diffusion tensor imaging (DTI) scans from a MR system using
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data…