Philips Medical Systems Nederland B.v. Azurion 7 M12 recall: Philips has identified instances where X-ray imaging may not be initiated
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1677-2026
- Published
- April 8, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems Nederland B.v.
- Product type
- Medical devices
- Units affected
- 1994 units (360 US, 1634 OUS)
- Distribution
- Nationwide
- Made in
- Netherlands
- Lot / code info
- (1) System Code: 722078; UDI: 00884838085251; System Serial Number: 302, 585, 410, 753, 523, 293, 840, 291, 536, 537, 870, 613, 598, 679, 224, 829, 249, 546, 478, 479, 513, 278, 437, 438, 456, 566, 833, 488, 327, 550, 571, 544, 545, 465, 376, 31, 921, 700, 165, 445, 539, 654, 656, 657, 658, 436, 1007, 862, 786, 308, 396, 576, 121, 239, 383, 491, 125, 500, 645, 379, 399, 803, 415, 296, 136, 890, 626, 697, 559, 532, 538, 185, 166, 257, 258, 684, 359, 143, 661, 824, 649, 493, 720, 520, 521, 556, 557, 558, 187, 902, 269, 317, 522, 611, 677, 709, 535, 170, 723, 251, 298, 22, 305, 725, 728, 524, 290, 441, 384, 480, 481, 364, 294, 906, 240, 256, 603, 289, 719, 163, 215, 271, 788, 412, 627, 141, 644, 292, 844, 876, 748, 483, 497, 309, 232, 425, 635, 612, 797, 247, 733, 162, 517, 888, 923, 204, 205, 206, 195, 197, 218, 444, 986, 561, 503, 689, 499, 135, 589, 38, 977, 738, 476, 380, 453, 24, 690, 356, 343, 344, 174, 898, 482, 378, 764, 693, 694, 770, 771, 772, 773, 439, 191, 938, 980, 981, 982, 307, 628, 235, 139, 178, 179, 265, 266, 511, 526, 746, 264, 287, 361, 362, 1003, 348, 897, 845, 960, 1036, 967, 374, 375, 564, 142, 469, 934, 565, 835, 836, 929, 975, 681, 1069, 1070, 1071, 1072, 893, 1061, 382, 324, 604, 713, 114, 849, 326, 659, 346, 347, 26, 621, 553, 325, 648, 1058, 1068, 339, 340, 414, 687, 688, 237, 958, 751, 881, 698, 610, 20, 846, 625, 424, 800, 838, 590, 150, 762, 485, 726, 727, 1039, 847, 959, 962, 865, 180, 371, 769, 570, 572, 1031, 792, 702, 196, 28, 29, 683, 541, 213, 320, 1025, 943, 651, 722, 242, 36, 431, 672, 177, 578, 912, 355, 504, 699, 270, 306, 190, 366, 52, 398, 470, 562, 489, 147, 160, 743, 152, 759, 630, 729, 151, 334, 815, 429, 449, 757, 506, 703, 149, 818, 647, 419, 328, 584, 223, 192, 779, 468, 222, 200, 134, 262, 855, 284, 918, 919, 277, 976, 13, 15, 502, 817, 795, 498, 574, 577, 254, 255, 448, 27, 631, 632, 599, 370, 216, 451, 428, 467, 450, 102, 662, 189, 633, 512, 175, 796, 459, 460, 718, 462, 464, 891, 768, 263, 944, 669, 112, 23, 452, 338, 367, 440, 490, 181, 145, 250, 705, 49, 50, 51, 617, 336, 226, 761, 220, 221, 1037, 799, 839, 354, 153, 154, 957, 471, 357, 130, 132, 133, 261, 695, 158, 507, 854, 477, 785, 789, 404, 877, 806, 422, 194, 801, 802, 140, 608, 878, 952, 10, 123, 319, 686, 805, 397, 127, 128, 253, 594, 988, 300, 301, 321, 332, 35, 17, 341, 563, 314, 316, 496, 641, 953, 30, 950, 331, 652, 1005, 349, 236, 979, 831, 501, 841, 842, 905, 682, 843, 830, 323, 315, 680, 365, 282, 283, 602, 203, 202, 40, 673, 674, 675, 676, 609, 295, 368, 369, 615, 618, 310, 199, 455, 668, 352, 193, 280, 667, 884, 567, 183, 749, 692, 32, 241, 935, 407, 408, 505, 138, 157, 752, 461, 463, 423, 554, 252, 169, 426, 650, 351, 442, 750, 968, 207, 208, 209, 210, 211, 622, 623, 629, 129, 529, 144, 400, 401, 411, 259, 457, 579, 926, 597, 899, 322, 978, 173, 418, 637, 670, 909, 601, 37, 1033, 1035, 227, 1009, 1008, 989, 984, 987, 907, 1081, 1004, 1086, 1084, 1085, 1048, 1059, 1034, 1080, 1015, 875, 1026, 1077, 716, 717, 1045, 940, 942, 945, 946, 947, 948, 1052, 971, 1062, 1063, 994, 701, 993, 620, 1075, 1046, 970, 985, 798, 1074, 856, 710, 1011, 1064, 1065, 1066, 1067, 1000, 1056, 992, 663, 1057, 964, 965, 1043, 735, 736, 819, 820, 272, 1049, 1073, 1079, 712, 714, 732, 774, 381, 233, 638, 318, 593, 624, 775, 776, 21, 267, 268, 487, 421, 760, 542, 614, 864, 148, 837, 458, 596, 377, 281, 784, 991, 528, 42, 44, 45, 41, 43, 466, 704, 219, 217, 260, 103, 430, 342, 126, 184, 778, 547, 586, 879, 159, 363, 896, 642, 472, 39, 172, 782, 961, 33, 643, 581, 552, 525, 311, 312, 510, 857, 731, 432, 433, 636, 954, 447, 333, 358, 540, 619, 492, 527, 600, 816, 434, 892, 990, 124, 122, 146, 201, 509, 495, 515, 353, 303, 304, 186, 286, 548, 518, 486, 474, 475, 1047, 853, 243, 244, 335, 549, 46, 616, 972, 533, 777, 182, 1032, 1051, 850, 696, 345, 575, 939, 869, 874, 871, 1010, 1029, 1030, 18, 551, 1013, 917, 543, 516, 12, 416, 155, 446, 685, 715, 225, 1082, 913, 582, 908, 19, 435, 16, 11, 167, 664, 901, 508, 530, 198; (2) System Code: 7222…
- UPC codes
- 00884838085251
Products covered by this recall
Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
Check your product
The notice names 9 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838085251
Serial number
On the rating plate, often next to the date of manufacture.
- 302
- 585
- 410
- 753
- 523
- 293
- 840
- 291
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems Nederland B.v. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 6, Best
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems Nederland B.v. medical devices recalled?
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838085251. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Philips Medical Systems Nederland B.v. Allura Xper FD. Model Numbers: 722026 recall: Firmware issue potentially leading to loss of geometry movements
Firmware issue potentially leading to loss of geometry movements.
-
Class II · 2 months ago
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232 recall: Two software related issues: 1) Potential image quality degradation
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate…
-
Class II · 2 months ago
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067 recall: Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural…
-
Class II · 2 months ago
Philips Allura, Cardio Vascular-Allura Centron recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Azurion under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Allura under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.