Philips Medical Systems Nederland B.v. Azurion 7 B20 recall: Philips has identified instances where X-ray imaging may not be initiated
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1676-2026
- Published
- April 8, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems Nederland B.v.
- Product type
- Medical devices
- Units affected
- 1523 units (322 US, 1201 OUS)
- Distribution
- Nationwide
- Made in
- Netherlands
- Lot / code info
- (1) System Code: 722068; UDI: 00884838085367; System Serial Number: 547, 178, 467, 335, 147, 205, 170, 455, 494, 564, 418, 95, 440, 247, 225, 64, 149, 42, 153, 324, 323, 320, 85, 89, 109, 407, 239, 505, 271, 443, 104, 370, 322, 136, 373, 442, 524, 308, 354, 75, 279, 400, 196, 510, 110, 446, 78, 330, 73, 288, 285, 295, 406, 528, 329, 184, 350, 462, 304, 457, 470, 472, 319, 520, 490, 243, 103, 479, 261, 262, 150, 277, 172, 117, 130, 356, 143, 146, 84, 251, 469, 431, 361, 76, 374, 270, 114, 511, 477, 232, 137, 282, 487, 413, 337, 101, 213, 82, 265, 275, 485, 459, 276, 62, 506, 608, 473, 417, 302, 314, 88, 126, 211, 468, 422, 454, 410, 283, 231, 548, 466, 495, 207, 167, 445, 537, 438, 173, 233, 289, 536, 192, 204, 77, 166, 313, 236, 133, 481, 280, 307, 577, 210, 258, 328, 325, 199, 480, 53, 72, 364, 491, 423, 509, 246, 614, 142, 376, 539, 257, 140, 369, 195, 500, 508, 97, 87, 303, 48, 44, 301, 550, 460, 118, 293, 551, 478, 522, 568, 448, 242, 238, 68, 160, 83, 544, 230, 434, 435, 360, 93, 36, 372, 179, 306, 182, 583, 327, 244, 382, 428, 432, 215, 447, 338, 399, 339, 316, 555, 552, 570, 347, 148, 134, 38, 560, 158, 535, 318, 518, 250, 336, 185, 206, 50, 489, 521, 37, 79, 529, 111, 254, 427, 394, 71, 47, 534, 292, 497, 591, 35, 227, 359, 527, 387, 168, 128, 129, 396, 515, 486, 543, 69, 453, 174, 483, 484, 334, 425, 181, 408, 415, 141, 365, 471, 540, 584, 553, 602, 240, 617, 604, 593, 589, 526, 559, 612, 450, 556, 557, 574, 603, 601, 609, 616, 597, 605, 575, 516, 624, 599, 563, 590, 198, 600, 228, 380, 420, 397, 212, 475, 180, 268, 269, 414, 74, 562, 115, 349, 291, 519, 234, 219, 532, 533, 60, 398, 618, 549, 46, 96, 416, 249, 531, 426, 162, 290, 346, 49, 326, 286, 125, 102, 113, 124, 138, 362, 127, 145, 299, 317, 571, 187, 183, 45, 331, 106, 465, 281, 112, 501, 393, 116, 273, 188, 264, 139, 503, 122, 439, 135, 248, 404, 390, 263, 90, 300, 224, 67, 61, 157, 154, 100, 217, 253, 391, 144, 379, 202, 105, 226, 315, 464, 197, 63, 252, 175, 274, 163, 186, 260, 502, 152, 498, 156, 621, 296, 208, 430, 161, 436, 409, 155, 298, 191, 278, 368, 94, 66, 108, 200, 201, 98, 189, 311, 367, 297, 132, 507, 451; (2) System Code: 722226; UDI: 00884838099272; System Serial Number: 120, 122, 123, 124, 125, 126, 127, 153, 67, 209, 271, 675, 112, 1052, 1091, 113, 121, 139, 158, 1078, 1145, 239, 105, 678, 114, 362, 758, 403, 702, 108, 545, 197, 380, 281, 688, 592, 593, 495, 319, 687, 921, 182, 183, 325, 690, 100, 272, 680, 724, 350, 891, 280, 270, 254, 107, 144, 761, 681, 691, 779, 294, 845, 559, 116, 343, 98, 301, 117, 876, 1021, 244, 332, 781, 782, 534, 341, 374, 78, 11, 441, 129, 180, 918, 442, 220, 555, 189, 190, 871, 446, 842, 255, 848, 208, 554, 955, 360, 337, 159, 400, 757, 326, 323, 175, 427, 464, 468, 778, 683, 143, 262, 482, 971, 391, 392, 173, 793, 147, 204, 467, 1017, 260, 381, 912, 410, 508, 419, 313, 472, 894, 318, 473, 383, 689, 973, 463, 823, 181, 205, 361, 59, 756, 379, 1008, 266, 1002, 398, 75, 786, 19, 1125, 213, 214, 215, 1007, 814, 969, 970, 469, 219, 462, 296, 149, 704, 573, 458, 987, 141, 711, 693, 510, 882, 226, 509, 584, 21, 1050, 647, 225, 201, 202, 303, 666, 184, 243, 178, 179, 250, 1139, 674, 1062, 575, 1121, 368, 950, 861, 176, 712, 829, 413, 997, 249, 589, 1046, 708, 738, 1083, 717, 697, 1126, 665, 416, 386, 764, 977, 595, 859, 1113, 721, 1043, 797, 957, 1006, 498, 60, 232, 423, 682, 789, 698, 331, 1097, 1094, 378, 808, 1107, 393, 949, 718, 663, 195, 411, 1092, 412, 715, 1143, 775, 1073, 982, 282, 785, 617, 958, 221, 222, 604, 430, 1027, 1067, 415, 810, 1114, 61, 1147, 314, 822, 561, 1026, 1102, 762, 306, 359, 978, 1060, 972, 83, 199, 1013, 920, 1136, 898, 835, 836, 795, 671, 799, 800, 899, 905, 838, 944, 1123, 335, 363, 1039, 731, 399, 869, 725, 701, 626, 967, 623, 744, 501, 953, 1018, 513, 533, 677, 449, 119, 367, 927, 480, 732, 974, 344, 824, 984, 487, 919, 605, 230, 523, 686, 327, 1138, 424, 684, 477, 1049, 298, 809, 10, 227, 520, 521, 522, 993, 874, 259, 140, 628, 530, 506, 909, 142, 733, 263, 490, 494, 633, 634, 396, 547, …
- UPC codes
- 00884838085367, 00884838099272
Products covered by this recall
Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
Check your product
The notice names 18 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838085367
- 00884838099272
Serial number
On the rating plate, often next to the date of manufacture.
- 547
- 178
- 467
- 335
- 147
- 205
- 170
- 455
- 120
- 122
- 123
- 124
- 125
- 126
- 127
- 153
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems Nederland B.v. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 6, Best
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems Nederland B.v. medical devices recalled?
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838085367, 00884838099272. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Philips Medical Systems Nederland B.v. Allura Xper FD. Model Numbers: 722026 recall: Firmware issue potentially leading to loss of geometry movements
Firmware issue potentially leading to loss of geometry movements.
-
Class II · 2 months ago
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232 recall: Two software related issues: 1) Potential image quality degradation
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate…
-
Class II · 2 months ago
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067 recall: Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural…
-
Class II · 2 months ago
Philips Allura, Cardio Vascular-Allura Centron recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Azurion under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Allura under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.